Scintillation imaging as a high-resolution, remote, versatile 2D detection system for MR-linac quality assurance.
Article in Medical physics, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
4 citing papers in PubMed.
- Assessment of scintillation and Cherenkov imaging as beam shape verification method in CyberKnife® radiotherapy.Journal of applied clinical medical physics · 2024Article
- Development of a three-dimensional scintillation detector for pencil beam verification in proton therapy patient-specific quality assurance.Medical physics · 2024Article
- Radioluminescence imaging feasibility for robotic radiosurgery field size quality assurance.Medical physics · 2022Article
- Visual Isocenter Position Enhanced Review (VIPER): a Cherenkov imaging-based solution for MR-linac daily QA.Medical physics · 2021Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
Abstract
purposeTo demonstrate the potential benefits of remote camera-based scintillation imaging for routine quality assurance (QA) measurements for magnetic resonance guided radiotherapy (MRgRT) linear accelerators.
methodsA wall-mounted CMOS camera with a time-synchronized intensifier was used to image photons produced from a scintillation screen in response to dose deposition from a 6 MV FFF x-ray beam produced by a 0.35 T MR-linac. The oblique angle of the field of view was corrected using a projective transform from a checkerboard calibration target. Output sensitivity and constancy was measured using the scintillator and benchmarked against an A28 ion chamber. Field cross-plane and in-plane profiles were measured for field sizes ranging from 1.68 × 1.66 cm
resultsThe scintillation imaging system showed comparable sensitivity and linearity to the ion chamber in response to changes in machine output down to 0.5 MU (R
conclusionsRemote scintillation imaging of a two-dimensional screen provided a rapid, versatile, MR-compatible solution to many routine quality assurance procedures including output constancy, profile flatness and symmetry constancy, MLC position verification and isocenter size. This method is high-resolution, does not require post-irradiation readout, and provides simple, instantaneous data acquisition. Full automation of the readout and processing could make this a very simple but effective QA tool, and is adaptable to all medical accelerators.