ArticleArthroplasty today2020
Taper fretting corrosion with Stryker Anato stem after hip replacement.
Article in Arthroplasty today, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed, 2 citations in OpenAlex.
- Femoral Stem Dislocation Caused by Trunnionosis Along with Adverse Local Tissue Reaction: A Case Report and a New Technique of Head to Cone Cementing.The archives of bone and joint surgery · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
We report a case of head-neck taper fretting corrosion in a patient who had a total hip replacement with a noncemented Stryker Anato femoral stem and a V40 metal head with a Stryker Tritanium hemispherical socket with a highly cross-linked polyethylene liner (metal on polyethylene) (Stryker, Mahwah, NJ, USA). A 57-year-old man presented with early-onset hip pain after right total hip arthroplasty. Workup was negative for infection. Metal artifact reduction sequence MRI revealed an encapsulated fluid mass. Metal ion cobalt level was elevated at 6 ppb. The patient underwent right revision total hip arthroplasty with excellent results at 1-year follow-up.