ArticleJournal of medical biochemistry2020
Glucose meters: What's the laboratory reference glucose?
Article in Journal of medical biochemistry, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed, 5 citations in OpenAlex.
- Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundThe accuracy of glucose meters is evaluated by comparing their results with those from a reference laboratory glucose analyser. The main scientific societies recommend the use of a prompt glycolysis inhibitor such as citrate for an accurate glucose determination. In the present preliminary study, we discuss the bias between capillary and plasma glucose measured concentrations, determined in two Italian clinical laboratories, using tubes containing an NaF and citrate mixture in liquid and granular form.
methods139 volunteers in whom 75 g OGTT was requested were recruited. Basal capillary glucose was determined using Abbott FreeStyle Precision Neo in Brescia (n=63), while clinical laboratory reference P-glucose was determined using tubes containing NaF/K
results96.82% and 97.37% of capillary determinations were within specifications when liquid and granular citrate mixture tubes were used, respectively. Conversely, only 73.02% and 80.26% of determinations were within criteria using NaF.
conclusionsIt's important to know what is the laboratory reference glucose in evaluating glucose meters' accuracy. The evaluation of glucometers' accuracy with respect to a reference laboratory may be wrong if tubes containing only NaF are used due to in vitro glycolysis. Only tubes containing citrate mixture permit the correct evaluation of glucose meters' accuracy.