Evidence map›Paper›PMID 32539300›Full record

ArticleAnnals of laboratory medicine2020

Stability of pH, Blood Gas Partial Pressure, Hemoglobin Oxygen Saturation Fraction, and Lactate Concentration.

Ariadna Arbiol-Roca, Claudia Elizabeth Imperiali, Dolors Dot-Bach, José Valero-Politi, Macarena Dastis-Arias

Abstract read
In one paragraph

Article in Annals of laboratory medicine, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.

0numbers the graph read from it
0cells of the map it votes in
8citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

8 citing papers in PubMed.

  1. Article
  2. Article
  3. Review
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
4 · The record

Corrections and comments

5 · Who and what money

Authors and funding

5 authors.

Ariadna Arbiol-RocaLaboratori Clínic Territorial Metropolitana Sud-Hospital Universitari de Bellvitge. Hospitalet de Llobregat, Barcelona, Spain.ORCID 0000-0002-5206-3818
Claudia Elizabeth ImperialiLaboratori Clínic Territorial Metropolitana Sud-Hospital Universitari de Bellvitge. Hospitalet de Llobregat, Barcelona, Spain.ORCID 0000-0002-7130-922X
Dolors Dot-BachLaboratori Clínic Territorial Metropolitana Sud-Hospital Universitari de Bellvitge. Hospitalet de Llobregat, Barcelona, Spain.ORCID 0000-0003-2642-8200
José Valero-PolitiLaboratori Clínic Territorial Metropolitana Sud-Hospital Universitari de Bellvitge. Hospitalet de Llobregat, Barcelona, Spain.ORCID 0000-0003-2949-000X
Macarena Dastis-AriasLaboratori Clínic Territorial Metropolitana Sud-Hospital Universitari de Bellvitge. Hospitalet de Llobregat, Barcelona, Spain.ORCID 0000-0003-3705-2739

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundThe storage temperature and time of blood gas samples collected in syringes constitute preanalytical variables that could affect blood gas or lactate concentration measurement results. We analyzed the effect of storage temperature and time delay on arterial or venous blood gas stability related to pH, partial pressure of carbon dioxide (pCO2) and oxygen (pO2), hemoglobin oxygen saturation (sO2), and lactate concentration.

methodsIn total, 1,200 arterial and venous blood sample syringes were analyzed within 10 minutes of collection. The samples were divided into different groups to determine parameter stability at 25, 4-8, and 0-3.9°C and at different storage times, 60, 45, 30, and 15 minutes. Independent sample groups were used for each analysis. Percentage deviations were calculated and compared with acceptance stability limits (1.65× coefficient of variation). Additionally, sample group sub analysis was performed to determine whether stability was concentration-dependent for each parameter.

resultsThe pH was stable over all storage times at 4-8 and 0-3.9°C and up to 30 minutes at 25°C. pCO2 was stable at ≤60 minutes at all temperatures. pO2 was stable for 45 minutes at 0-3.9°C, and sO2 was stable for 15 minutes at 25°C and for ≤60 minutes at 0-3.9°C. Lactate concentration was stable for 45 minutes at 0-3.9°C. Subanalysis showed that stability was concentration-dependent.

conclusionsThe strictest storage temperature and time criteria (0-3.9°C, 45 minutes) should be adopted for measuring pH, pCO2, pO2, sO2, and lactate concentration in blood gas syringes.

Indexed as

Blood Specimen CollectionGasesHemoglobinsHumansHydrogen-Ion ConcentrationLactic AcidOxygenPartial PressureTemperatureTime FactorsGasesHemoglobinsLactic AcidOxygenBlood gasesPreanalytical variablesStabilityStorageTemperatureTime

Identifiers

PMID32539300
PMCPMC7295962

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.