ArticleInternational ophthalmology2020
Stability of a non-applanating handheld liquid patient interface for femtosecond laser-assisted cataract surgery.
Article in International ophthalmology, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Capsulotomy morphology in cataract surgery: comparison of five femtosecond lasers.Journal of cataract and refractive surgery · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeA stable and reliable vacuum is crucial for the correct planning and performance of femtosecond laser-assisted cataract surgery (FLACS) in order to avoid complications such as suction loss and cyclorotation. This study investigates, for the first time, the impact of different vacuum levels on the stability of the application of a liquid patient interface for FLACS in view of break-away forces.
methodsBreak-away forces were measured using a multifunctional material testing machine with a mounted digital manometer. Sixteen porcine eyes were docked to the patient interface of a femtosecond laser platform (FEMTO LDV Z8), and the impact of different vacuum levels between 300 and 500 mbar investigated.
resultsMean break-away forces for each vacuum level were as follows: 1.78 N (± 0.58 N) for 300 mbar; 2.24 N (± 0.68 N) for 350 mbar; 2.66 N (± 0.68 N) for 400 mbar; 2.86 N (± 0.77 N) for 420 mbar; and 3.49 N (± 0.86 N) for 500 mbar.
conclusionThe stability increases with the vacuum in a nearly linear manner. Vacuum levels higher than 500 mmHg and lower than 350 mmHg are not recommended for FLACS.