Evidence map›Paper›PMID 32481729›Full record

ArticleSensors (Basel, Switzerland)2020

Development of an Accurate Resonant Frequency Controlled Wire Ultrasound Surgical Instrument.

Jungsuk Kim, Kyeongjin Kim, Sun-Ho Choe, Hojong Choi

Open access · goldAbstract read
In one paragraph

Article in Sensors (Basel, Switzerland), 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 11 papers.

0numbers the graph read from it
0cells of the map it votes in
11citing papers in PubMed
1.7field-weighted citation impact, top 13% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

11 citing papers in PubMed, 16 citations in OpenAlex.

  1. Article
  2. Article
  3. Fisheye lens design for solar-powered mobile ultrasound devices.Technology and health care : official journal of the European Society for Engineering and Medicine · 2022
    Article
  4. Novel dual-resistor-diode limiter circuit structures for high-voltage reliable ultrasound receiver systems.Technology and health care : official journal of the European Society for Engineering and Medicine · 2022
    Article
  5. Article
  6. Article
  7. Article
  8. Development of a low-cost six-axis alignment instrument for flexible 2D and 3D ultrasonic probes.Technology and health care : official journal of the European Society for Engineering and Medicine · 2021
    Article
  9. Article
  10. Article
  11. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors at 2 institutions in 1 country.

Jungsuk KimDepartment of Biomedical Engineering, Gachon University, 534-2, Hambakmoe-ro, Incheon 21936, Korea.ORCID 0000-0001-5866-465X
Kyeongjin KimDepartment of Medical IT Convergence Engineering, Kumoh National Institute of Technology, 350-27 Gumi-daero, Gumi 39253, Korea.
Sun-Ho ChoeR&D Center, Metabiomed Corporation, 215 Osongsaenmyeong1-ro, Chenongu 28161, Korea.
Hojong ChoiDepartment of Medical IT Convergence Engineering, Kumoh National Institute of Technology, 350-27 Gumi-daero, Gumi 39253, Korea.
Kumoh National Institute of Technology · KRGachon University · KR

Funding

MOTIE S2520804National Research Foundation of Korea 2017M3A9E2056461National Research Foundation of Korea 2020R1A2C4001606
6 · The paper itself

Abstract

Our developed wire ultrasound surgical instrument comprises a bolt-clamped Langevin ultrasonic transducer (BLUT) fabricated by PMN-PZT single crystal material due to high mechanical quality factor and electromechanical coupling coefficient, a waveguide in the handheld instrument, and a generator instrument. To ensure high performance of wire ultrasound surgical instruments, the BLUT should vibrate at an accurate frequency because the BLUT's frequency influences hemostasis and the effects of incisions on blood vessels and tissues. Therefore, we implemented a BLUT with a waveguide in the handheld instrument using a developed assembly jig process with impedance and network analyzers that can accurately control the compression force using a digital torque wrench. A generator instrument having a main control circuit with a low error rate, that is, an output frequency error rate within ±0.5% and an output voltage error rate within ±1.6%, was developed to generate the accurate frequency of the BLUT in the handheld instrument. In addition, a matching circuit between the BLUT and generator instrument with a network analyzer was developed to transfer displacement vibration efficiently from the handheld instrument to the end of the waveguide. Using the matching circuit, the measured S-parameter value of the generator instrument using a network analyzer was -24.3 dB at the resonant frequency. Thus, our proposed scheme can improve the vibration amplitude and accuracy of frequency control of the wire ultrasound surgical instrument due to developed PMN-PZT material and assembly jig process.

Indexed as

Surgical InstrumentsTransducersEquipment DesignUltrasonicsbolt-clamped Langevin ultrasonic transducergenerator instrumenthandheld instrumentwire ultrasound surgical instrument

Identifiers

PMID32481729
PMCPMC7309115
OpenAlexW3031940715

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.