ArticleJournal of neuro-oncology2020
Assessment of unintended shifts during frame-based stereotactic radiosurgery using cone beam computed tomography image guidance.
Article in Journal of neuro-oncology, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed, 1 citations in OpenAlex.
- Optimizing Gamma Knife radiosurgery: Analysis of gains achieved through upgradation of technology.Journal of radiosurgery and SBRT · 2025Article
- Discrepancies between frame- and CBCT-based stereotactic space definition on the Gamma Knife Icon.Journal of applied clinical medical physics · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
10 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo investigate the frequency, magnitude and possible causes of frame-shifts that may occur between treatment planning and treatment delivery when performing Gamma Knife radiosurgery with rigid frame-based immobilization.
methodsDifferences between computed tomography (CT) framed fiducial stereotactic coordinate reference and cone beam computed tomography stereotactic coordinates after image registration were recorded for 49 frame-based GK radiosurgery cases performed using the Gamma Knife Icon. Parameters recorded include rotational shifts, translational shifts, and the GK-computed Maximum Shot Displacement (MSD) between the two stereotactic coordinate spaces. Other patient-specific parameters were collected and linear regression analysis was performed to evaluate predictors of increased displacement.
resultsThe median values of rotational shifts were: pitch 0.14°, yaw 0.17°, and roll 0.13°. The median absolute values of translational shifts were: left-right 0.39 mm, anteroposterior 0.14 mm, and superior-inferior 0. 22 mm. The median value of MSD was 0.71 mm. Twelve cases (24.5%) had a MSD of greater than 1.0 mm. Male gender was associated with increased MSD (p = 0.013) and translational shifts (root-mean-squared value, p = 0.017). Cases with large differences between right and left sided pin lengths were also associated with increased MSD (p = 0.011).
conclusionsThe use of CBCT image guidance in frame-based GK radiosurgery allows unintended frame shifts to be identified and corrected. A significant fraction (24.5%) of patients had large enough shifts to result in a MSD of greater than 1.0 mm. Male gender and eccentrically placed frames were associated with increased MSD, and particular care should be taken in these cases.