ArticlePloS one2020
Differentiation between mechanically loose and fixed press-fit implants using quantitative acoustics and load self-referencing: A phantom study on shoulder prostheses in polyurethane foam.
Article in PloS one, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Optical phantoms for the characterization and validation of imaging and spectroscopy: a review.Biophotonics discovery · 2026Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
This study proposes to use cross-interface quantitative acoustics (ci-qA) and load self-referencing (LSR) to assess implant stability in a radiation-free, inexpensive, rapid, and quantitative manner. Eight bone analog specimens, made from polyurethane foam, were implanted with a cementless stemless shoulder implant-first in a fixed and later in a loose configuration-and measured using ci-qA under two load conditions. The loose implants exhibited higher micromotion and lower pull-out strength than their stable counterparts, with all values falling within the range of reported reference values. All acoustic characteristics differentiated between loose and fixed implants (maximum area-under-curve AUC = 1.0 for mean total signal energy, AUC = 1.0 for mean total signal energy ratio, AUC = 0.8 for harmonic ratio, and AUC = 0.92 for load self-referencing coefficient). While these results on bone substitute material will need to be confirmed on real bone specimen, ci-qA could ultimately facilitate the assessment of primary stability during implantation surgery and avoid unnecessary revision through quantitative evaluation of secondary stability during follow-up.