ArticleMedical physics2020
Dose calculation errors as a component of failing IROC lung and spine phantom irradiations.
Article in Medical physics, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
4 citing papers in PubMed, 12 citations in OpenAlex.
- Latent density discrepancies in commercial lung-equivalent inserts and their clinical dosimetric impact.Journal of applied clinical medical physics · 2026Article
- Comprehensive end-to-end dosimetry audit for stereotactic body radiotherapy in spine, lung, and soft tissue.Journal of applied clinical medical physics · 2025Article
- Early results of a remote dosimetry audit program for lung stereotactic body radiation therapy.Physics and imaging in radiation oncology · 2024Article
- Photon beam modeling variations predict errors in IMRT dosimetry audits.Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors at 2 institutions in 1 country.
Funding
Abstract
purposeBetween July 2013 and August 2019, 22% of the imaging and radiation oncology core (IROC) spine, and 15% of the moving lung phantom irradiations have failed to meet established acceptability criteria. The spine phantom simulates a highly modulated stereotactic body radiation therapy (SBRT) case, whereas the lung phantom represents a low-to-none modulation moving target case. In this study, we assessed the contribution of dose calculation errors to these phantom results and evaluated their effects on failure rates.
methodsWe evaluated dose calculation errors by comparing the calculation accuracy of various institutions' treatment planning systems (TPSs) vs IROC-Houston's previously established independent dose recalculation system (DRS). Each calculation was compared with the measured dose actually delivered to the phantom; cases in which the recalculation was more accurate were interpreted as a deficiency in the institution's TPS. A total of 258 phantom irradiation plans (172 lung and 86 spine) were recomputed.
resultsOverall, the DRS performed better than the TPSs in 47% of the spine phantom cases. However, the DRS was more accurate in 93% of failing spine phantom cases (with an average improvement of 2.35%), indicating a deficiency in the institution's treatment planning system. Deficiencies in dose calculation accounted for 60% of the overall discrepancy between measured and planned doses among spine phantoms. In contrast, lung phantom DRS calculations were more accurate in only 35% and 42% of all and failing lung phantom cases respectively, indicating that dose calculation errors were not substantially present. These errors accounted for only 30% of the overall discrepancy between measured and planned doses.
conclusionsDose calculation errors are common and substantial in IROC spine phantom irradiations, highlighting a major failure mode in this phantom and in clinical treatment management of these cases. In contrast, dose calculation accuracy had only a minimal contribution to failing lung phantom results, indicating that other failure modes drive problems with this phantom and similar clinical treatments.