ArticleMedical physics2020
X-ray-induced acoustic computed tomography for guiding prone stereotactic partial breast irradiation: a simulation study.
Article in Medical physics, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 10 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
10 citing papers in PubMed.
- XACT-guided interventional procedures: A feasibility study.Medical physics · 2026Article
- Seeing the invisible: In vivo dosimetry in radiotherapy with radiacoustic IMAGING.Medical physics · 2025Review
- 4DPhysics in medicine and biology · 2024Article
- Discrete Wavelet Transformation for the Sensitive Detection of Ultrashort Radiation Pulse with Radiation-induced Acoustics.IEEE transactions on radiation and plasma medical sciences · 2024Article
- X-ray-induced acoustic computed tomography and its applications in biomedicine.Journal of biomedical optics · 2024Review
- Model-Based 3-D X-Ray Induced Acoustic Computerized Tomography.IEEE transactions on radiation and plasma medical sciences · 2023Article
- Ring artifacts removal in X-ray-induced acoustic computed tomography.Journal of innovative optical health sciences · 2022Article
- Model-Based X-Ray-Induced Acoustic Computed Tomography.IEEE transactions on ultrasonics, ferroelectrics, and frequency control · 2021Article
- Synchrotron X-ray induced acoustic imaging.Scientific reports · 2021Article
- Review of in vivo optical molecular imaging and sensing from x-ray excitation.Journal of biomedical optics · 2021Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
Abstract
purposeThe aim of this study is to investigate the feasibility of x-ray-induced acoustic computed tomography (XACT) as an image guidance tool for tracking x-ray beam location and monitoring radiation dose delivered to the patient during stereotactic partial breast irradiation (SPBI).
methodsAn in-house simulation workflow was developed to assess the ability of XACT to act as an in vivo dosimetry tool for SPBI. To evaluate this simulation workflow, a three-dimensional (3D) digital breast phantom was created by a series of two-dimensional (2D) breast CT slices from a patient. Three different tissue types (skin, adipose tissue, and glandular tissue) were segmented and the postlumpectomy seroma was simulated inside the digital breast phantom. A treatment plan was made with three beam angles to deliver radiation dose to the seroma in breast to simulate SPBI. The three beam angles for 2D simulations were 17°, 90° and 159° (couch angles were 0 degrees) while the angles were 90 degrees (couch angles were 0°, 27°, 90°) in 3D simulation. A multi-step simulation platform capable of modelling XACT was developed. First, the dose distribution was converted to an initial pressure distribution. The propagation of this pressure disturbance in the form of induced acoustic waves was then modeled using the k-wave MATLAB toolbox. The waves were then detected by a hemispherical-shaped ultrasound transducer array (6320 transducer locations distributed on the surface of the breast). Finally, the time-varying pressure signals detected at each transducer location were used to reconstruct an image of the initial pressure distribution using a 3D time-reversal reconstruction algorithm. Finally, the reconstructed XACT images of the radiation beams were overlaid onto the structure breast CT.
resultsIt was found that XACT was able to reconstruct the dose distribution of SPBI in 3D. In the reconstructed 3D volumetric dose distribution, the average doses in the GTV (Gross Target Volume) and PTV (Planning Target Volume) were 86.15% and 80.89%, respectively. When compared to the treatment plan, the XACT reconstructed dose distribution in the GTV and PTV had a RMSE (root mean square error) of 2.408 % and 2.299 % over all pixels. The 3D breast XACT imaging reconstruction with time-reversal reconstruction algorithm can be finished within several minutes.
conclusionsThis work explores the feasibility of using the novel imaging modality of XACT as an in vivo dosimeter for SPBI radiotherapy. It shows that XACT imaging can provide the x-ray beam location and dose information in deep tissue during the treatment in real time in 3D. This study lays the groundwork for a variety of future studies related to the use of XACT as a dosimeter at different cancer sites.