ArticleBMJ paediatrics open2020
Evaluation of a continuous neonatal temperature monitor for low-resource settings: a device feasibility pilot study.
Article in BMJ paediatrics open, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to 2 registered trials, which are not on this map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Evaluating the Accuracy of a Neonatal Temperature Monitor in Comparison With Existing Temperature Monitors
Pilot Clinical Assessment of Low-cost Infant Incubator in Monitoring Temperature and Treating Hypothermia in Infants
Who cites it
2 citing papers in PubMed, 4 citations in OpenAlex.
- Premature Infants With Pulmonary Complications Exhibit Decreased Skin Maturity Early After Birth: A Cross-Sectional Study.Pediatric pulmonology · 2026Article
- Acceptability and feasibility of the BEMPU TempWatch for hypothermia monitoring in neonatal care in Ghana: a qualitative study of clinicians' perspectives.BMJ public health · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors at 2 institutions in 2 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectiveEvaluate a novel continuous temperature monitor in a low-resource neonatal ward.
designWe developed a low-cost continuous neonatal temperature monitor (NTM) for use in low-resource settings. Accuracy of NTM was initially assessed in the laboratory. Clinical evaluation then was performed in a neonatal ward in a central hospital in Malawi; eligible neonates (<1 week of age) were recruited for continuous temperature monitoring with NTM and a Philips Intellivue MP30 Patient Monitor. INTERVENTIONS AND OUTCOME MEASURES: The temperature probes of NTM and the reference patient monitor were attached to the infant's abdomen, and core temperature was continuously recorded for up to 3 hours. Axillary temperatures were taken every hour. We compared temperatures measured using NTM, the patient monitor and the axillary thermometer.
resultsLaboratory temperature measurements obtained with NTM were within 0.059°C (range: -0.035°C to 0.195°C) of a reference thermometer. A total of 39 patients were recruited to participate in the clinical evaluation of NTM; data from four patients were excluded due to faulty hardware connections. The mean difference in measured temperatures between the NTM and the Intellivue MP30 was -0.04°C (95% CI -0.52°C to 0.44°C).
conclusionNTM meets ISO 80601-2-56 standards for accuracy and is an appropriate, low-cost continuous temperature monitor for neonatal wards in low-resource settings. TRIAL REGISTRATION NUMBERS: NCT03965312 and NCT03866122.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials