Evidence map›Paper›PMID 32420677›Full record

ArticleJournal of veterinary internal medicine2020

Evaluation of the Freestyle Optium Neo H point-of-care device for measuring blood glucose concentrations in sick calves.

Tolga Karapinar, Kenan Cagri Tumer, Sébastien Buczinski

Open access · goldAbstract read
In one paragraph

Article in Journal of veterinary internal medicine, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.

0numbers the graph read from it
0cells of the map it votes in
5citing papers in PubMed
0.7field-weighted citation impact, top 30% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

5 citing papers in PubMed, 10 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors at 2 institutions in 2 countries.

Tolga KarapinarDepartment of Internal Medicine, Faculty of Veterinary Medicine, Firat University, Elazig, Turkey.ORCID https://orcid.org/0000-0003-1724-491X
Kenan Cagri TumerDepartment of Internal Medicine, Faculty of Veterinary Medicine, Firat University, Elazig, Turkey.
Sébastien BuczinskiDépartement des sciences cliniques, Faculté de médecine vétérinaire, Université de Montréal, Saint-Hyacinthe, Québec, Canada.ORCID https://orcid.org/0000-0002-8460-4885
Fırat University · TRCegep de Saint Hyacinthe · CA

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundData on the performance of a glucometer in calves with different diseases are currently lacking.

objectiveThe primary objective of this study was to evaluate the reliability of a point of care glucometer in calves affected by different diseases relative to a traditional bench-top autoanalyzer. ANIMALS: One hundred ninety-six calves with different disorders in a referral hospital.

methodsProspective study. Venous blood samples were used for the determination of glucose concentrations in blood and plasma using the Freestyle Optium Neo H and autoanalyzer, respectively. Data were subjected to Passing-Bablok regression and Bland-Altman plots. The Freestyle Optium Neo H was the test method and the autoanalyzer was the reference method. The diagnostic performance of the glucometer relative to the autoanalyzer was assessed using 3 different plasma glucose concentrations.

resultsThe Passing-Bablok regression for the glucometer against the reference method revealed the presence of both proportional bias (1.12; 95% confidence interval [CI], 1.07-1.18) and constant bias (-11.25; 95% CI, -16.0 to -7.70). The glucometer yielded 92.2%-100% sensitivity and 86.4%-96% specificity for the assessing glucose concentration based on different concentration thresholds. CONCLUSIONS AND CLINICAL IMPORTANCE: The Freestyle Optium Neo H showed proportional and constant biases relative to the reference method. The glucometer showed poor performance according to criteria recommended by the International Standards Organization and the American Society for Veterinary Clinical Pathology. However, the glucometer determined hypoglycemia with high sensitivity and specificity therefore it might be used to diagnose hypoglycemia in calves with different diseases until calf-specific POC glucometers are developed.

Indexed as

Point-of-Care SystemsAnimalsAnimals, NewbornBlood Chemical AnalysisBlood GlucoseCattleCattle DiseasesHypoglycemiaProspective StudiesSensitivity and SpecificityBlood Glucoseaccuracybovineglucometerprecisionsensitivityspecificity

Identifiers

PMID32420677
PMCPMC7379022
OpenAlexW3026810513

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked