Evidence map›Paper›PMID 32385798›Full record

ArticleJournal of robotic surgery2021

Patient injuries and malfunctions associated with robotic prostatectomy: review of the manufacturer and user facility device experience database.

Hudson Pierce, Brittany Roberts, Douglas Scherr, Christopher Barbieri, Jennifer Anger, Timothy McClure, Bilal Chughtai

Abstract read
PubMed Publisher
In one paragraph

Article in Journal of robotic surgery, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.

0numbers the graph read from it
0cells of the map it votes in
7citing papers in PubMed
0.6field-weighted citation impact, top 32% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

7 citing papers in PubMed, 10 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 4 institutions in 1 country.

Hudson PierceDepartment of Urology, Weill Cornell Medical College-New York Presbyterian, 425 East 61st Street, 12th Floor, New York, NY, 10065, USA.
Brittany RobertsDepartment of Obstetrics and Gynecology, University of South Florida, Tampa, FL, 33602, USA.
Douglas ScherrDepartment of Urology, Weill Cornell Medical College-New York Presbyterian, 425 East 61st Street, 12th Floor, New York, NY, 10065, USA.
Christopher BarbieriDepartment of Urology, Weill Cornell Medical College-New York Presbyterian, 425 East 61st Street, 12th Floor, New York, NY, 10065, USA.
Jennifer AngerDivision of Urology, Department of Surgery, Cedars-Sinai Medical Center, Los Angeles, CA, USA.
Timothy McClureDepartment of Urology, Weill Cornell Medical College-New York Presbyterian, 425 East 61st Street, 12th Floor, New York, NY, 10065, USA.
Bilal ChughtaiDepartment of Urology, Weill Cornell Medical College-New York Presbyterian, 425 East 61st Street, 12th Floor, New York, NY, 10065, USA. bic9008@med.cornell.edu.ORCID http://orcid.org/0000-0002-0515-2578
Cornell University · USCedars-Sinai Medical Center · USNewYork–Presbyterian Hospital · USUniversity of South Florida · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Robotic prostatectomy is the most commonly performed robotic procedure in the United States. Increasing utilization of this procedure necessitates characterization of robot malfunctions and associated patient injuries. We performed a review of adverse events reported to a publicly available database. We searched the Manufacturer and User Facility Device Experience (MAUDE) database for reported adverse events (RAE) involving the intuitive surgical system. Reports involving prostatectomy from 2014 to 2019 were extracted and analyzed for data regarding death, patient injury, and device malfunction. Of 9109 reported adverse events (RAE), 602 were extracted for robotic prostatectomy over the study period. Seven were patient deaths (1.2%), 53 (8.8%) were patient injuries (Table 1), and 542 (90.0%) were malfunctions (Table 2). Malfunctions resulted in 25 aborted cases, 21 open conversions, and 25 laparoscopic conversions (71/542, 13.1%; Fig. 1). Instrument failures comprised the majority (76.4%) of malfunctions. Seven malfunctions (1.3%) resulted in patient injury. The most common device-related injury involved the monopolar curved scissors. No reported deaths were related to robot malfunction. Instrument failures comprise majority of the malfunctions of the Da Vinci robot during robot-assisted laparoscopic prostatectomy. When malfunctions do occur they are usually recoverable and rarely lead to patient injury.

Indexed as

Data AnalysisDatabases, FactualEquipment FailureHumansIntraoperative ComplicationsLaparoscopyMaleProstatectomyRobotic Surgical ProceduresTime FactorsWounds and Injuriesda VinciDevice malfunctionInjuryMAUDERobotic prostatectomy

Identifiers

PMID32385798
OpenAlexW3021108308

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.