Evidence map›Paper›PMID 32377319›Full record

ArticleThe journal of advanced prosthodontics2020

Does the prosthesis weight matter? 3D finite element analysis of a fixed implant-supported prosthesis at different weights and implant numbers.

João Paulo Mendes Tribst, Amanda Maria de Oliveira Dal Piva, Alexandre Luiz Souto Borges, Vinicius Anéas Rodrigues, Marco Antonio Bottino, Cornelis Johannes Kleverlaan

Abstract read
In one paragraph

Article in The journal of advanced prosthodontics, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.

0numbers the graph read from it
0cells of the map it votes in
9citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

9 citing papers in PubMed.

  1. Article
  2. Review
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

João Paulo Mendes TribstDepartment of Dental Materials and Prosthodontics, São Paulo State University (UNESP), Institute of Science and Technology, São José dos Campos, SP, Brazil.ORCID https://orcid.org/0000-0002-5412-3546
Amanda Maria de Oliveira Dal PivaDepartment of Dental Materials and Prosthodontics, São Paulo State University (UNESP), Institute of Science and Technology, São José dos Campos, SP, Brazil.ORCID https://orcid.org/0000-0002-3844-2053
Alexandre Luiz Souto BorgesDepartment of Dental Materials and Prosthodontics, São Paulo State University (UNESP), Institute of Science and Technology, São José dos Campos, SP, Brazil.ORCID https://orcid.org/0000-0002-5707-7565
Vinicius Anéas RodriguesDepartment of Dentistry, Faculty of Pindamonhangaba (FUNVIC), Pindamonhangaba, SP, Brazil.ORCID https://orcid.org/0000-0002-7209-868X
Marco Antonio BottinoDepartment of Dental Materials and Prosthodontics, São Paulo State University (UNESP), Institute of Science and Technology, São José dos Campos, SP, Brazil.ORCID https://orcid.org/0000-0003-0077-3161
Cornelis Johannes KleverlaanDepartment of Dental Materials Science, Academic Centre for Dentistry Amsterdam (ACTA), The University of Amsterdam and Free University Amsterdam, Amsterdam, Netherlands.ORCID https://orcid.org/0000-0002-3218-8031

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeThis study evaluated the influence of prosthesis weight and number of implants on the bone tissue microstrain. MATERIALS AND

methodsFifteen (15) fixed full-arch implant-supported prosthesis designs were created using a modeling software with different numbers of implants (4, 6, or 8) and prosthesis weights (10, 15, 20, 40, or 60 g). Each solid was imported to the computer aided engineering software and tetrahedral elements formed the mesh. The material properties were assigned to each solid with isotropic and homogeneous behavior. The friction coefficient was set as 0.3 between all the metallic interfaces, 0.65 for the cortical bone-implant interface, and 0.77 for the cancellous bone-implant interface. The standard earth gravity was defined along the Z-axis and the bone was fixed. The resulting equivalent strain was assumed as failure criteria.

resultsThe prosthesis weight was related to the bone strain. The more implants installed, the less the amount of strain generated in the bone. The most critical situation was the use of a 60 g prosthesis supported by 4 implants with the largest calculated magnitude of 39.9 mm/mm, thereby suggesting that there was no group able to induce bone remodeling simply due to the prosthesis weight.

conclusionHeavier prostheses under the effect of gravity force are related to more strain being generated around the implants. Installing more implants to support the prosthesis enables attenuating the effects observed in the bone. The simulated prostheses were not able to generate harmful values of peri-implant bone strain.

Indexed as

BiomechanicsBone tissueDental implantsFinite element analysisProsthodontics

Identifiers

PMID32377319
PMCPMC7183854

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.