ArticleObstetrics and gynecology2020
Temporal Trends of Urogynecologic Mesh Reports to the U.S. Food and Drug Administration.
Article in Obstetrics and gynecology, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT06292702 (Reduce Voiding Dysfunction After Laparoscopic Burch Colposuspension by Using Lateral Tension for the Anterior Vaginal Wall Before Cooper's Ligament Suspension in Stress Incontinence Women.), which is not on this map. Cited by 7 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Reduce Voiding Dysfunction After Laparoscopic Burch Colposuspension by Using Lateral Tension for the Anterior Vaginal Wall Before Cooper's Ligament Suspension in Stress Incontinence Women. (A Randomized Controlled Clinical Trial).
Who cites it
7 citing papers in PubMed, 13 citations in OpenAlex.
- Uterine-Sparing vNOTES vs. laparoscopic lateral suspension for pelvic organ prolapse repair: a prospective comparative study of feasibility and early outcomes.Frontiers in surgery · 2026Article
- Article
- New native tissue repair for pelvic organ prolapse: Medium-term outcomes of laparoscopic vaginal stump-round ligament fixation.World journal of clinical cases · 2023Article
- Advances in vaginal bioengineering: Applications, techniques, and needs.Current research in physiology · 2023Review
- Knowledge mapping and visualization analysis of pelvic organ prolapse repair with mesh from 2001 to 2021.Frontiers in bioengineering and biotechnology · 2023Article
- Complications Reported to the Food and Drug Administration: A Cross-sectional Comparison of Urogynecologic Meshes.Female pelvic medicine & reconstructive surgery · 2022Article
- Symptomatic improvement after mesh removal: a prospective longitudinal study of women with urogynaecological mesh complications.BJOG : an international journal of obstetrics and gynaecology · 2021Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 1 institution in 2 countries.
Funding
Abstract
objectiveTo compare temporal trends of urogynecologic mesh medical device reports with sentinel U.S. Food and Drug Administration (FDA) notices and to examine all linked reports of patient death.
methodsThe Reed Tech Navigator is an online tool used to extract and analyze data in the Manufacturer and User Facility Device Experience database. We used FDA product codes to search for reports of synthetic mesh for stress urinary incontinence (SUI) and pelvic organ prolapse (POP). Reports with "death" listed as a patient event were queried further.
resultsThere were 43,970 medical device reports involving urogynecologic mesh reported to the FDA between August 2000 and January 2019, with most occurring after the 2011 FDA communication (n=43,018, 97.8%). Of these medical device reports, 64.6% (n=28,422) were for SUI products, 27.0% (n=11,876) were for transvaginal POP products, and 8.4% (n=3,672) were for transabdominal POP products. Peak reporting occurred in 2013, which corresponded with the first major plaintiff verdicts and class action vaginal mesh settlement. There were 645 (1.5%) medical device reports of death, with 49.3% (n=318) having no cause of death identified. Only four (0.6%) could be directly attributed to the initial surgery.
conclusionSince 2011, urogynecologic mesh has been the focus of thousands of medical device reports annually, although the data submitted are often incomplete. Few reported deaths are directly attributable to mesh products, and the majority lack sufficient information to draw causal conclusions. A high-quality registry is necessary to improve our understanding of the patient effect from urogynecologic mesh products.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.