Observational studyNeuromodulation : journal of the International Neuromodulation Society2020

Catheter Access Port (Computed Tomography) Myelography in Intrathecal Drug Delivery Troubleshooting: A Case Series of 70 Procedures.

Elmar M Delhaas, Biswadjiet S Harhangi, Sander P G Frankema, Frank J P M Huygen, Aad van der Lugt

Open access · bronzeAbstract readObservational Study
In one paragraph

Observational study in Neuromodulation : journal of the International Neuromodulation Society, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
0.9field-weighted citation impact, top 27% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 9 citations in OpenAlex.

  1. Review
  2. Article
  3. Review
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 1 institution in 1 country.

Elmar M DelhaasCenter for Pain Medicine, Erasmus MC University Medical Center Rotterdam, Rotterdam, The Netherlands.ORCID https://orcid.org/0000-0003-1342-1261
Biswadjiet S HarhangiDepartment of Neurosurgery, Erasmus MC University Medical Center Rotterdam, Rotterdam, The Netherlands.ORCID https://orcid.org/0000-0001-6983-3164
Sander P G FrankemaCenter for Pain Medicine, Erasmus MC University Medical Center Rotterdam, Rotterdam, The Netherlands.
Frank J P M HuygenCenter for Pain Medicine, Erasmus MC University Medical Center Rotterdam, Rotterdam, The Netherlands.
Aad van der LugtDepartment of Radiology and Nuclear Medicine, Erasmus MC University Medical Center Rotterdam, Rotterdam, The Netherlands.
Erasmus MC · NL

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectivesIntrathecal drug delivery is used for the treatment of intractable spasticity, dystonia, and pain. When the symptomatology fails to respond to therapy, the cause could be failure of the medication infusion. The purpose of this study is to assess pump catheter access port (CAP)-myelography and CAP-CT-myelography as advanced imaging methods in treatment failure. MATERIALS AND

methodsWe analyzed observational routinely collected data of 70 CAP procedures with 2D/3D reconstructions and additional imaging of 53 adult patients where the cause of treatment was unclear between November 2013 and November 2018. CAP-myelography and CAP-CT myelography were performed with postprocessing 2D/3D reconstructions. When myelography could not be obtained or when the result did not reveal the cause of the treatment failure, additional procedures, such as noncontrast CT, MRI, lumbar puncture CT, and 111Indium-DTPA SPECT-CT, were performed.

resultsCAP fluid aspiration prior to contrast medium injection was not possible (N = 17). In one case, contrast was injected into the pump pocket unintentionally (N = 1). Of 70 procedures, 24% were unaspiratable. The remaining CAP myelography examinations (N = 52) had limited value for the diagnosis. CAP-CT myelography (N = 50) was normal (N = 31). The abnormal results (N = 19) were dorsal dural leak (N = 5), subdural catheter position (N = 2), limited rostral flow of contrast material (N = 4), limited and abnormal contrast distribution (N = 3), obstruction of rostral flow (N = 2), a leak at the pump-catheter connection (N = 1), and a sheared catheter localized in the pump pocket (N = 2). Limited contrast distributions were found to be false positive findings (N = 2). Four normal CT-CAP myelographic procedures were false negatives, as the reference tests revealed a cause of intrathecal drug delivery (ITDD) failure. The CAP-CT procedures resulted in a sensitivity of 81% (17/21) and a specificity of 93% (27/29).

conclusionsCAP-CT myelography with 2D/3D reconstructions is an essential step in the diagnostic algorithm for cases involving ITDD failure.

Indexed as

Drug Delivery SystemsInjections, SpinalMyelographyAdultCathetersHumansRetrospective StudiesTomography, X-Ray ComputedCatheter access portcomputed tomographyintrathecal therapy failuremyelographyprogrammable pump

Identifiers

PMID32267597
PMCPMC7687167
OpenAlexW3014265915

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.