ArticleMedical physics2020
Effects of kV, filtration, dose, and object size on soft tissue and iodine contrast in dedicated breast CT.
Article in Medical physics, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 10 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
10 citing papers in PubMed, 20 citations in OpenAlex.
- Hybrid simulation of breast CT for assessing microcalcification detectability.Journal of medical imaging (Bellingham, Wash.) · 2025Article
- Soft-tissue lesion and microcalcification detectability in cone-beam breast CT: cascaded system analysis.Journal of medical imaging (Bellingham, Wash.) · 2025Article
- Assessment of pulmonary vascular anatomy: comparing augmented reality by holograms versus standard CT images/reconstructions using surgical findings as reference standard.European radiology experimental · 2024Article
- Maximizing microcalcification detectability in low-dose dedicated cone-beam breast CT: parallel cascades-based theoretical analysis.Journal of medical imaging (Bellingham, Wash.) · 2024Article
- Discrimination tasks in simulated low-dose CT noise.Medical physics · 2023Article
- Cone-beam breast CT using an offset detector: effect of detector offset and image reconstruction algorithm.Physics in medicine and biology · 2022Article
- Dose reduction potential in diagnostic single energy CT through patient-specific prefilters and a wider range of tube voltages.Medical physics · 2022Article
- High-resolutionJournal of medical imaging (Bellingham, Wash.) · 2021Article
- Computer-aided diagnosis of masses in breast computed tomography imaging: deep learning model with combined handcrafted and convolutional radiomic features.Journal of medical imaging (Bellingham, Wash.) · 2021Article
- Sparse-view, short-scan, dedicated cone-beam breast computed tomography: image quality assessment.Biomedical physics & engineering express · 2020Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 2 institutions in 1 country.
Funding
Abstract
purposeClinical use of dedicated breast computed tomography (bCT) requires relatively short scan times necessitating systems with high frame rates. This in turn impacts the x-ray tube operating range. We characterize the effects of tube voltage, beam filtration, dose, and object size on contrast and noise properties related to soft tissue and iodine contrast agents as a way to optimize imaging protocols for soft tissue and iodine contrast at high frame rates.
methodsThis study design uses the signal-difference-to-noise ratio (SDNR), noise-equivalent quanta (NEQ), and detectability (d´) as measures of imaging performance for a prototype breast CT scanner that utilizes a pulsed x-ray tube (with a 4 ms pulse width) at 43.5 fps acquisition rate. We assess a range of kV, filtration, breast phantom size, and mean glandular dose (MGD). Performance measures are estimated from images of adipose-equivalent breast phantoms machined to have a representative size and shape of small, medium, and large breasts. Water (glandular tissue equivalent) and iodine contrast (5 mg/ml) were used to fill two cylindrical wells in the phantoms.
resultsAir kerma levels required for obtaining an MGD of 6 mGy ranged from 7.1 to 9.1 mGy and are reported across all kV, filtration, and breast phantom sizes. However, at 50 kV, the thick filters (0.3 mm of Cu or Gd) exceeded the maximum available mA of the x-ray generator, and hence, these conditions were excluded from subsequent analysis. There was a strong positive association between measurements of SDNR and d' (R
conclusionsThe choice of 60 kV with 0.2 mm Gd filtration provides a good trade-off for maximizing both soft tissue and iodine contrast. This scanning technique takes advantage of the ~50 keV Gd k-edge to produce contrast and can be achieved within operating range of the x-ray generator used in this work. Imaging at 60 kV allows for a greater range in dose delivered to the large breast sizes when uniform image quality is desired across all breast sizes. While imaging performance metrics (i.e., detectability index and SDNR) were shown to be strongly correlated, the methodologies presented in this work for the estimation of NEQ (and subsequently d') provides a meaningful description of the spatial resolution and noise characteristics of this prototype bCT system across a range of beam quality, dose, and object sizes.