Evidence map›Paper›PMID 32218581›Full record

ArticlePloS one2020

Simplified pulse wave velocity measurement in children: Is the pOpmètre valid?

Saïd Bichali, Alexandra Bruel, Marion Boivin, Gwénaëlle Roussey, Bénédicte Romefort, Jean-Christophe Rozé, Emma Allain-Launay

Registry-linked trialAbstract read
In one paragraph

Article in PloS one, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT02991703 (Simplified Pulse Wave Velocity Measurement, Validation Study of the pOpmètre in Children), which is not on this map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

NCT02991703 nacompletednot on this map

Simplified Pulse Wave Velocity Measurement, Validation Study of the pOpmètre in Children

TypeinterventionalSponsorNantes University HospitalRan2016 to 2017Enrolled60ConditionsCardiovascular Diseases, Arteriosclerosis, AtherosclerosisArmspOpmètre®, Sphygmocor
3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Article
  2. Review
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Saïd BichaliPediatric Cardiology Unit, Nantes University Hospital, Nantes, France.ORCID 0000-0003-2220-6061
Alexandra BruelPediatric Nephrology Unit, Nantes University Hospital, Nantes, France.
Marion BoivinClinical Investigation Center CIC 004, INSERM-Nantes, Nantes University Hospital, Nantes, France.
Gwénaëlle RousseyPediatric Nephrology Unit, Nantes University Hospital, Nantes, France.
Bénédicte RomefortPediatric Cardiology Unit, Nantes University Hospital, Nantes, France.
Jean-Christophe RozéDepartment of Neonatal Intensive Care, Nantes University Hospital, Nantes, France.ORCID 0000-0003-3367-512X
Emma Allain-LaunayPediatric Nephrology Unit, Nantes University Hospital, Nantes, France.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

In population exposed to cardiovascular risk, aortic stiffness is an important marker which is assessed by carotid-to-femoral pulse wave velocity (PWV). In childhood, the validated applanation tonometer SphygmoCor® can be used to measure PWV, but is limited in routine practice by the child's cooperation and operator's experience. An alternative device, the pOpmètre® is validated in adults and rapidly measures finger-to-toe PWV using 2 oxymeter-like sensors. The aim of this study is to validate the pOpmètre® device in children aged between 4 and 8 years. We compared simultaneous PWV measurements of the two devices, SphygmoCor® and pOpmètre®, in a training group, using the Bland-Altman method. Then we proposed an algorithm to correct pOpmètre® PWV (PWVpop). Finally, we validated this new algorithm in a validation group of children using the Bland-Altman method. This prospective study enrolled 26 children in the training group. Mean PWVpop was 3.919 ± 0.587 m/s and mean SphygmoCor® PWV was 4.280 ± 0.383 m/s, with a difference of -0.362(CI95%(-0.546;-0.178)) m/s. A new algorithm was defined using transit time (TTpop): corrected PWVpop (m/s) = 0.150/TTpop(s) + 1.381*Height(m) + 1.148. We enrolled 24 children in the validation group. Mean corrected PWVpop was 4.231 ± 0.189 m/s and mean SphygmoCor® PWV was 4.208 ± 0.296 m/s with a corrected difference of 0.023(CI95%(-0.086;0.131)) m/s. With this algorithm correction, we found an agreement between PWV measured by the SphygmoCor® and the pOpmètre®, with a difference of less than 10%. Using this algorithm, the pOpmètre® could be used in clinical or research practice in young children exposed to cardiovascular risk. (This study was registered as NCT02991703).

Indexed as

ChildChild, PreschoolFemaleHumansMalePulse Wave Analysis

Identifiers

PMID32218581
PMCPMC7100956

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.