Evidence map›Paper›PMID 32167702›Full record

ArticleInternational braz j urol : official journal of the Brazilian Society of Urology

Comparison of automated irrigation systems using an in vitro ureteroscopy model.

Donald Fedrigon, Luay Alshara, Manoj Monga

Abstract read
In one paragraph

Article in International braz j urol : official journal of the Brazilian Society of Urology. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.

0numbers the graph read from it
0cells of the map it votes in
9citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

9 citing papers in PubMed.

  1. Review
  2. Article
  3. Article
  4. Article
  5. Review
  6. Article
  7. Article
  8. Editorial Comment: Comparison of automated irrigation systems using an in vitro ureteroscopy model.International braz j urol : official journal of the Brazilian Society of Urology
    Article
  9. Pediatric Urology highlighted.International braz j urol : official journal of the Brazilian Society of Urology
    Article
4 · The record

Corrections and comments

5 · Who and what money

Authors and funding

3 authors.

Donald FedrigonCleveland Clinic, Glickman Kidney & Urological Institute, Cleveland, OH, USA.
Luay AlsharaCleveland Clinic, Glickman Kidney & Urological Institute, Cleveland, OH, USA.
Manoj MongaCleveland Clinic, Glickman Kidney & Urological Institute, Cleveland, OH, USA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

introductionTwo automated irrigation systems have been released for use during endoscopic procedures such as ureteroscopy: the Cogentix RocaFlow® (CRF) and Thermedx FluidSmart® (TFS). Accurate pressure control using automated systems may help providers maintain irrigation pressures within a safe range while also providing clear visualization. Our objective was to directly compare these systems based on their pressure accuracy, pressure-flow relationships, and fluid heating capabilities in order to help providers better utilize the temperature and pressure settings of each system. MATERIALS AND

methodsAn in vitro ureteroscopy model was used for testing, consisting of a short semirigid ureteroscope (6/7, 5F, 31cm Wolf 425612) connected to a continuous digital pressure transducer (Meriam m1550). Each system pressure output and flow-rate, via 100mL beaker filling time, was measured using multiple trials at pressure settings between 30 and 300mmHg. Output fluid temperature was monitored using a digital thermometer (Omega DP25-TH).

resultsThe pressure output of both systems exceeded the desired setting across the entire tested range, a difference of 15.7±2.4mmHg for the TFS compared to 5.2±1.5mmHg for the CRF (p < 0.0001). Related to this finding, the TFS also had slightly higher flow rates across all trials (7±2mL/min). Temperature testing revealed a similar maximum temperature of 34.0⁰C with both systems, however, the TFS peaked after only 8 minutes and started to plateau as early as 4-5 minutes into the test, while the CRF took over 18 minutes to reach a similar peak.

conclusionsOur in vitro ureteroscopy testing found that the CRF system had better pressure accuracy than the TFS system but with noticeably slower fluid heating capabilities. Each system provided steady irrigation at safe pressures within their expected operating parameters with small differences in performance that should not limit their ability to provide steady irrigation at safe pressures.

Indexed as

Therapeutic IrrigationUreteroscopyPressureTemperatureUreteroscopesResearchTechnologyUreteroscopy

Identifiers

PMID32167702
PMCPMC7088507

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked