Evidence map›Paper›PMID 32056504›Full record

ArticlePerfusion2020

The effect of flow and pressure on the intraoxygenator flow path of different contemporary oxygenators: an in vitro trial.

Rik Hj Hendrix, Alexander Ks Yeung, Yuri M Ganushchak, Patrick W Weerwind

Abstract read
In one paragraph

Article in Perfusion, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Rik Hj HendrixDepartment of Extra-Corporeal Circulation, Maastricht University Medical Centre, Maastricht, The Netherlands.ORCID 0000-0001-6733-6287
Alexander Ks YeungDepartment of Extra-Corporeal Circulation, Maastricht University Medical Centre, Maastricht, The Netherlands.
Yuri M GanushchakDepartment of Extra-Corporeal Circulation, Maastricht University Medical Centre, Maastricht, The Netherlands.
Patrick W WeerwindDepartment of Extra-Corporeal Circulation, Maastricht University Medical Centre, Maastricht, The Netherlands.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

introductionThis study analyzed the effect of different flows and pressures on the intraoxygenator flow path in three contemporary oxygenators and its consequences for oxygen transfer efficiency.

methodsIn an experimental setup, intraoxygenator flow path parameters were analyzed at post-oxygenator pressures of 150, 200, and 250 mm Hg and at flows ranging from 2 L/min to the oxygenators' maximum permitted flow, with and without pulsatility. The oxygen gradient and the oxygen transfer per minute and per 100 mL blood were calculated using previously collected clinical data and compared with the flow path parameters.

resultsIncreasing pressure did not affect the flow path parameters, whereas pulsatile flow led to significantly increased dynamic oxygenator blood volumes. Increased flow resulted in decreased values of the flow path parameters in all oxygenators, indicating increased flow through short pathways in the oxygenator. In parallel, oxygen transfer/100 mL blood decreased in all oxygenators (average 2.5 ± 0.4 to 2.4 ± 0.3 mL/dL, p > 0.001) and the oxygen gradient increased from 229 ± 45 to 287 ± 29 mm Hg, p > 0.001, indicating decreased oxygen transfer efficiency. Oxygen transfer/min increased (101 ± 15 to 143 ± 20 mL/min/m

conclusionVarying trans-membrane oxygenator pressures did not lead to changes in the intraoxygenator flow path, while an increased flow exhibited lower flow path parameters resulting in less efficient use of the gas exchange compartment. The latter was confirmed by a decrease in O

Indexed as

PressureHumansIn Vitro TechniquesOxygenatorsPulsatile Flowflow pathoxygenatoroxygenator blood volumeoxygenator efficiency

Identifiers

PMID32056504
PMCPMC7536508

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.