ArticleJournal of biomedical physics & engineering2019
Validation of the Gel & Wax Boluses and Comparison of their Dosimetric Performance with Virtual Bolus.
Article in Journal of biomedical physics & engineering, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Implementation of Customized Homogenizing Devices in Radiotherapy for the Treatment of Non-Melanoma Skin Cancer: Three Case Reports.Reports (MDPI) · 2025Article
- Evaluation of 3D-printed bolus for radiotherapy using megavoltage X-ray beams.Radiological physics and technology · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundIn general, radiotherapy treatment planning is performed using the virtual bolus. It is necessary to investigate physical bolus in comparison to virtual one.
objectivesIn the present study, first, radiological properties of superflab Gel bolus and Paraffin wax bolus was investigated in terms of their relative electron density. Then, dosimetric performance of both the bolus (i.e. Gel and Parafin wax) was compared with Virtual bolus. MATERIAL AND
methodsIn This experimental study, the radiological property of Wax and Gel boluses was investigated using two methods. In one, the relative electron density of both the Gel and Wax boluses was calculated by measuring their linear attenuation coefficient where in another method relative electron density was calculated by recording their CT No directly from their CT scan. Later CT scan of solid water slab phantom (dimension 30x30x15 cm
resultsThe measured (calculated) relative electron density of wax and Gel bolus was found to be 0.958 (0.926) and 0.923 (0.907), respectively. Variation in dosimetric performance of Gel and Wax with reference to Virtual bolus was studied. However, on average, Gel bolus was more consistent with virtual bolus.
conclusionTo avoid any dose difference between, delivered (using physical bolus) and planned (using virtual bolus), the physical boluses should be investigated for their dosimetric performance in comparison to virtual bolus. The results obtained and methodology used in this study can be applied in routine radiotherapy practices.
Indexed as