ArticleJournal of applied clinical medical physics2020
A pitfall of using the circular-edge technique with image averaging for spatial resolution measurement in iteratively reconstructed CT images.
Article in Journal of applied clinical medical physics, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
4 citing papers in PubMed.
- Assessment of image quality and detectability in computed tomography: Inter-laboratory comparison of Fourier-based metrics.Journal of applied clinical medical physics · 2026Article
- Reduction of metal artefacts from bilateral hip prostheses during lower extremity computed tomography angiography: an experimental phantom study.Journal of medical radiation sciences · 2024Article
- Analysis of the precision of modulation transfer function measurements with a circular edge: influence of object circularity.Journal of medical imaging (Bellingham, Wash.) · 2023Article
- Simulation Analysis on the Performance of a Circular-Edge Technique in Measurements of the Modulation Transfer Function.Journal of medical physicsArticle
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
2 authors.
Funding
Abstract
The circular-edge technique using a low-contrast cylindrical object is commonly used to measure the modulation transfer functions (MTFs) in computed tomography (CT) images reconstructed with iterative reconstruction (IR) algorithms. This method generally entails averaging multiple images of the cylinder to reduce the image noise. We suspected that the cylinder edge shape depicted in the IR images might exhibit slight deformation with respect to the true shape because of the intrinsic nonlinearity of IR algorithms. Image averaging can reduce the image noise, but does not effectively improve the deformation of the edge shape; thereby causing errors in the MTF measurements. We address this issue and propose a method to correct the MTF. We scanned a phantom including cylindrical objects with a CT scanner (Ingenuity Elite, Philips Healthcare). We obtained cylinder images with iterative model reconstruction (IMR) algorithms. The images suggested that the depicted edge shape deforms and fluctuates depending on slice positions. Because of this deformation, image averaging can potentially cause additional blurring. We define the deformation function D that describes the additional blurring, and obtain D by analyzing multiple images. The MTF measured by the circular-edge method (referred to as MTF') can be thought of as the multiplication of the true MTF by the Fourier transformation (FT) of D. We thus obtain the corrected MTF (MTF