Low-dose megavoltage cone-beam computed tomography using a novel multi-layer imager (MLI).
Article in Medical physics, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
4 citing papers in PubMed.
- Impact of a novel multilayer imager on metal artifacts in MV-CBCT.Physics in medicine and biology · 2023Article
- Abbreviated on-treatment CBCT using roughness penalized mono-energization of kV-MV data and a multi-layer MV imager.Physics in medicine and biology · 2021Article
- Clinical translation of a new flat-panel detector for beam's-eye-view imaging.Physics in medicine and biology · 2020Article
- A rapid, accurate image simulation strategy for mega-voltage cone-beam computed tomography.Physics in medicine and biology · 2020Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
12 authors.
Funding
Abstract
purposeThe feasibility of low-dose megavoltage cone-beam acquisition (MVCBCT) using a novel, high detective quantum efficiency (DQE) multi-layer imager (MLI) was investigated. The aim of this work was to reconstruct MVCBCT images using the MLI at different total dose levels, and assess Hounsfield Unit (HU) accuracy, noise and contrast-to-noise ratio (CNR) for low-dose megavoltage cone-beam acquisition.
methodsThe MLI has four stacked layers; each layer contains a combination of copper filter/converter, gadolinium oxysulfide (GOS) scintillator and a-Si detector array. In total, 720 projections of a CATPHAN
resultsAverage HU uniformity for the MLI reconstructions was within a range of 95%-99% for all of the energies studied. Relative electron density estimation from HU values was within 0.4% ± 1.8% from nominal values. The CNR for MVCBCT based on MLI projections was 2-4× greater than from AS1200 projections. The 2.5 MV beam acquisition with the MLI exhibited the lowest noise and the best balance between CNR and dose for low-dose reconstructions.
conclusionsMegavoltage cone-beam acquisition imaging with a novel MLI prototype mounted on a clinical linear accelerator demonstrated substantial improvement over the standard AS1200 EPID. Further optimization of MVCBCT reconstruction, particularly for 2.5 MV acquisitions, will improve image metrics. Overall, the MLI improves CNR at substantially lower doses than currently required by conventional detectors. This new high DQE detector could provide high-quality MVCBCT at clinically acceptable doses.