ArticleIEEE transactions on bio-medical engineering2020
Joint Optimization of Collimator and Reconstruction Parameters in X-Ray Fluorescence Computed Tomography Using Analytical Point Spread Function and Model Observer.
Article in IEEE transactions on bio-medical engineering, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed, 4 citations in OpenAlex.
- Monte Carlo study of x-ray detection configurations for benchtop x-ray fluorescence computed tomography of gold nanoparticle-loaded objects.Physics in medicine and biology · 2020Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 3 institutions in 1 country.
Funding
Abstract
objectiveTo jointly optimize collimator design and image reconstruction algorithm in X-ray Fluorescence Computed Tomography (XFCT) for imaging low concentrations of high atomic number (Z) elements in small animal models.
methodsSingle pinhole (SPH) collimator and three types of multi-pinhole (MPH) collimators were evaluated. MPH collimators with 5, 7, and 9 pinholes using lead, stainless steel and brass were considered. A digital cylindrical phantom (0.5 mm
resultsA stainless steel focusing type MPH with 7 pinholes and 20 iterations of ML-EM provided the highest AUC.
conclusionMPH collimators outperformed SPH collimators for XFCT and consistently high AUCs were observed with focusing type MPH designs with 7 pinholes. SIGNIFICANCE: The combinations of collimator design and image reconstruction parameters that maximized AUC were identified, which could improve the performance of XFCT.