ArticleMedical physics2020

A mobile isocentric C-arm for intraoperative cone-beam CT: Technical assessment of dose and 3D imaging performance.

N M Sheth, T De Silva, A Uneri, M Ketcha, R Han, R Vijayan, G M Osgood, J H Siewerdsen

Open access · greenAbstract read
In one paragraph

Article in Medical physics, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 17 papers.

0numbers the graph read from it
0cells of the map it votes in
17citing papers in PubMed
5.4field-weighted citation impact, top 4% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

17 citing papers in PubMed, 34 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Review
  7. Article
  8. Article
  9. Article
  10. First clinical experience with a novel, mobile cone-beam CT system for treatment quality assurance in brachytherapy.Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al] · 2022
    Article
  11. Article
  12. Article
  13. Article
  14. Article
  15. Article
  16. Article
  17. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 1 institution in 1 country.

N M ShethDepartment of Biomedical Engineering, Johns Hopkins University, Baltimore, MD, USA.
T De SilvaDepartment of Biomedical Engineering, Johns Hopkins University, Baltimore, MD, USA.
A UneriDepartment of Biomedical Engineering, Johns Hopkins University, Baltimore, MD, USA.
M KetchaDepartment of Biomedical Engineering, Johns Hopkins University, Baltimore, MD, USA.
R HanDepartment of Biomedical Engineering, Johns Hopkins University, Baltimore, MD, USA.
R VijayanDepartment of Biomedical Engineering, Johns Hopkins University, Baltimore, MD, USA.
G M OsgoodDepartment of Orthopaedic Surgery, Johns Hopkins Medical Institutions, Baltimore, MD, USA.
J H SiewerdsenDepartment of Biomedical Engineering, Johns Hopkins University, Baltimore, MD, USA.
Johns Hopkins University · US

Funding

BIOMEDICAL ENGINEERING TRAINING PROGRAMT32GM007057 · NIGMS · JOHNS HOPKINS UNIVERSITY · PI KANOLD, PATRICK O, KARCHIN, RACHEL · 1985 to 2023
$10.4M
Intraoperative Imaging for Guidance, Patient Safety, and OR Quality AssuranceR01EB017226 · NIBIB · JOHNS HOPKINS UNIVERSITY · PI SIEWERDSEN, JEFFREY H · 2014 to 2017
$2.1M
NIBIB NIH HHS R01 EB017226NIGMS NIH HHS T32 GM007057NIH HHS -NIH HHS 017226NIH HHS EBNIH HHS R01Siemens Healthineers
6 · The paper itself

Abstract

purposeTo characterize the radiation dose and three-dimensional (3D) imaging performance of a recently developed mobile, isocentric C-arm equipped with a flat-panel detector (FPD) for intraoperative cone-beam computed tomography (CBCT) (Cios Spin 3D, Siemens Healthineers) and to identify potential improvements in 3D imaging protocols for pertinent imaging tasks.

methodsThe C-arm features a 30 × 30 cm

resultsGeometric calibration showed a high degree of reproducibility despite ~19 mm gantry flex over a nominal semicircular orbit. The dose for a CBCT scan varied from ~0.8-4.7 mGy for head protocols to ~6-38 mGy for body protocols. The MTF was consistent with sub-mm spatial resolution, with f

conclusionThe studies provided objective assessment of the dose and 3D imaging performance of a new C-arm, offering an important basis for clinical deployment and a benchmark for quality assurance. Modifications to standard 3D imaging protocols were identified that may improve performance or reduce radiation dose for pertinent imaging tasks.

Indexed as

Imaging, Three-DimensionalRadiation DosageCone-Beam Computed TomographyFluoroscopyHumansIntraoperative PeriodPhantoms, ImagingC-armcone-beam CTimage-guided surgeryimage qualityradiation dose

Identifiers

PMID31863480
PMCPMC8630541
OpenAlexW2996518595

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.