Evidence map›Paper›PMID 31841537›Full record

Trial reportPloS one2019

Accuracy of the Cosmed K5 portable calorimeter.

Scott E Crouter, Samuel R LaMunion, Paul R Hibbing, Andrew S Kaplan, David R Bassett

Abstract readRandomized Controlled Trial
In one paragraph

Trial report in PloS one, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 32 papers.

0numbers the graph read from it
0cells of the map it votes in
32citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

32 citing papers in PubMed.

  1. Trial
  2. Trial
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Pilot evaluation of a dance-based physical activity program for adults with Down syndrome.Translational journal of the American College of Sports Medicine · 2026
    Article
  10. Article
  11. Article
  12. Article
  13. Article
  14. Article
  15. Article
  16. Article
  17. Article
  18. Article
  19. Human-exoskeleton interaction portrait.Journal of neuroengineering and rehabilitation · 2024
    Article
  20. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Scott E CrouterDepartment of Kinesiology, Recreation, and Sport Studies, The University of Tennessee Knoxville, Knoxville, TN, United States of America.ORCID 0000-0003-1297-9538
Samuel R LaMunionDepartment of Kinesiology, Recreation, and Sport Studies, The University of Tennessee Knoxville, Knoxville, TN, United States of America.
Paul R HibbingDepartment of Kinesiology, Recreation, and Sport Studies, The University of Tennessee Knoxville, Knoxville, TN, United States of America.
Andrew S KaplanDepartment of Kinesiology, Recreation, and Sport Studies, The University of Tennessee Knoxville, Knoxville, TN, United States of America.
David R BassettDepartment of Kinesiology, Recreation, and Sport Studies, The University of Tennessee Knoxville, Knoxville, TN, United States of America.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeThe purpose of this study was to assess the accuracy of the Cosmed K5 portable metabolic system dynamic mixing chamber (MC) and breath-by-breath (BxB) modes against the criterion Douglas bag (DB) method.

methodsFifteen participants (mean age±SD, 30.6±7.4 yrs) had their metabolic variables measured at rest and during cycling at 50, 100, 150, 200, and 250W. During each stage, participants were connected to the first respiratory gas collection method (randomized) for the first four minutes to reach steady state, followed by 3-min (or 5-min for DB) collection periods for the resting condition, and 2-min collection periods for all cycling intensities. Collection periods for the second and third methods were preceded by a washout of 1-3 min. Repeated measures ANOVAs were used to compare metabolic variables measured by each method, for seated rest and each cycling work rate.

resultsFor ventilation (VE) and oxygen uptake (VO2), the K5 MC and BxB modes were within 2.1 l/min (VE) and 0.08 l/min (VO2) of the DB (p≥0.05). Compared to DB values, carbon dioxide production (VCO2) was significantly underestimated by the K5 BxB mode at work rates ≥150W by 0.12-0.31 l/min (p<0.05). K5 MC and BxB respiratory exchange ratio values were significantly lower than DB at cycling work rates ≥100W by 0.03-0.08 (p<0.05).

conclusionCompared to the DB method, the K5 MC and BxB modes are acceptable for measuring VE and VO2 across a wide range of cycling intensities. Both K5 modes provided comparable values to each other.

Indexed as

AdultBicyclingCalorimetryEquipment DesignExerciseExercise TestFemaleHumansMaleMobile ApplicationsMonitoring, AmbulatoryOxygen ConsumptionPulmonary Gas ExchangeReproducibility of ResultsRespirationRespiratory Function Tests

Identifiers

PMID31841537
PMCPMC6913985

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.