ArticleSkeletal radiology2020
Inter- and intra-reader reproducibility of shear wave elastography measurements for musculoskeletal soft tissue masses.
Article in Skeletal radiology, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed, 1 synthesis or guideline pooled it, 10 citations in OpenAlex.
- Diagnostic performance and clinical utility of shear-wave elastography in musculoskeletal soft-tissue tumors: a systematic review.Frontiers in oncology · 2026Pooled it
- Effect of knee and hip joint positions on passive stiffness of the rectus femoris and vastus lateralis in healthy individuals.Scientific reports · 2026Trial
- Ultrasound shear wave elastography for the assessment of pelvic floor muscle stiffness: technical and analytical challenges in practice.Ultrasonography (Seoul, Korea) · 2026Article
- Shear Wave Elastography in Musculoskeletal Imaging: A Narrative Review.Journal of clinical medicine · 2026Review
- Agreement of shear wave and strain elastography according to the size of the region of interest: a study of epidermal cysts.Ultrasonography (Seoul, Korea) · 2025Article
- Longitudinal and transverse muscle stiffness change differently with knee osteoarthritis and do not align with stiffness sensation.Frontiers in physiology · 2025Article
- Article
- Is thoracolumbar fascia shear-wave modulus affected by active and passive knee flexion?Journal of anatomy · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 2 institutions in 2 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectiveTo determine inter- and intra-reader reproducibility of shear wave elastography measurements for musculoskeletal soft tissue masses. MATERIALS AND
methodsIn all, 64 patients with musculoskeletal soft tissue masses were scanned by two readers prior to biopsy; each taking five measurements of shear wave velocity (m/s) and stiffness (kPa). A single lesion per patient was scanned in transverse and cranio-caudal planes. Depth measurements (cm) and volume (cm
resultsOf the 64 lesions scanned, 24 (38%) were malignant. Bland-Altman plots demonstrated negligible bias with wide LOA for all measurements. Transverse velocity was the most reliable measure-intraclass correlation (95% CI) = 0.917 (0.886, 1)-though reader 1 measures could be between 38% lower and 57% higher than reader 2 [ratio-scale bias (95% LOA) = 0.99 (0.64, 1.55)]. Repeatability coefficients indicated most disagreement resulted from poor within-reader reproducibility. LOA between readers calculated from means of five repeated measurements were narrower-transverse velocity ratio-scale bias (95% LOA) = 1.00 (0.74, 1.35). Depth affected both estimated velocity and repeatability; volume also affected repeatability.
conclusionThis study found poor repeatability of measurements with wide LOA due mostly to intra-reader variability. Transverse velocity was the most reliable measure; variability may be affected by lesion depth. At least five measurements should be reported with LOA to assist future comparability between shear wave elastography systems in evaluating soft tissue masses.