A novel approach to Verify air gap and SSD for proton radiotherapy using surface imaging.
Article in Radiation oncology (London, England), 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- Millimeter wave-based patient setup verification and motion tracking during radiotherapy.Medical physics · 2024Article
- Advances and potential of optical surface imaging in radiotherapy.Physics in medicine and biology · 2022Review
- Recent advanced in Surface Guided Radiation Therapy.Radiation oncology (London, England) · 2020Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors.
Funding
Abstract
purposeTo develop a novel approach to accurately verify patient set up in proton radiotherapy, especially for the verification of the nozzle - body surface air gap and source-to-skin distance (SSD), the consistency and accuracy of which is extremely important in proton treatment.
methodsPatient body surfaces can be captured and monitored with the optical surface imaging system during radiation treatment for improved intrafraction accuracy. An in-house software package was developed to reconstruct the patient body surface in the treatment position from the optical surface imaging reference capture and to calculate the corresponding nozzle - body surface air gap and SSD. To validate this method, a mannequin was scanned on a CT simulator and proton plans were generated for a Mevion S250 Proton machine with 20 gantry/couch angle combinations, as well as two different snout sizes, in the Varian Eclipse Treatment Planning Systems (TPS). The surface generated in the TPS from the CT scan was imported into the optical imaging system as an RT Structure for the purpose of validating and establishing a benchmark for ground truth comparison. The optical imaging surface reference capture was acquired at the treatment setup position after orthogonal kV imaging to confirm the positioning. The air gaps and SSDs calculated with the developed method from the surface captured at the treatment setup position (VRT surface) and the CT based surface imported from the TPS were compared to those calculated in TPS. The same approach was also applied to 14 clinical treatment fields for 10 patients to further validate the methodology.
resultsThe air gaps and SSDs calculated from our program agreed well with the corresponding values derived from the TPS. For the phantom results, using the CT surface, the absolute differences in the air gap were 0.45 mm ± 0.33 mm for the small snout, and 0.51 mm ± 0.49 mm for the large snout, and the absolute differences in SSD were 0.68 mm ± 0.42 mm regardless of snout size. Using the VRT surface, the absolute differences in air gap were 1.17 mm ± 1.17 mm and 2.1 mm ± 3.09 mm for the small and large snouts, respectively, and the absolute differences in SSD were 0.81 mm ± 0.45 mm. Similarly, for patient data, using the CT surface, the absolute differences in air gap were 0.42 mm ± 0.49 mm, and the absolute differences in SSD were 1.92 mm ± 1.4 mm. Using the VRT surface, the absolute differences in the air gap were 2.35 mm ± 2.3 mm, and the absolute differences in SSD were 2.7 mm ± 2.17 mm.
conclusionThese results showed the feasibility and robustness of using an optical surface imaging approach to conveniently determine the air gap and SSD in proton treatment, providing an accurate and efficient way to confirm the target depth at treatment.