Evidence map›Paper›PMID 31728679›Full record

ArticleThe international journal of cardiovascular imaging2020

Impact of baseline calibration on semiquantitative assessment of myocardial perfusion reserve by adenosine stress MRI.

Andreas Seitz, Giancarlo Pirozzolo, Udo Sechtem, Raffi Bekeredjian, Peter Ong, Heiko Mahrholdt

Abstract readComparative Study
PubMed Publisher
In one paragraph

Article in The international journal of cardiovascular imaging, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Andreas SeitzDepartment of Cardiology, Robert-Bosch-Krankenhaus, Auerbachstr. 110, 70376, Stuttgart, Germany. andreas.seitz@rbk.de.
Giancarlo PirozzoloDepartment of Cardiology, Robert-Bosch-Krankenhaus, Auerbachstr. 110, 70376, Stuttgart, Germany.
Udo SechtemDepartment of Cardiology, Robert-Bosch-Krankenhaus, Auerbachstr. 110, 70376, Stuttgart, Germany.
Raffi BekeredjianDepartment of Cardiology, Robert-Bosch-Krankenhaus, Auerbachstr. 110, 70376, Stuttgart, Germany.
Peter OngDepartment of Cardiology, Robert-Bosch-Krankenhaus, Auerbachstr. 110, 70376, Stuttgart, Germany.
Heiko MahrholdtDepartment of Cardiology, Robert-Bosch-Krankenhaus, Auerbachstr. 110, 70376, Stuttgart, Germany.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

In this study, we sought to investigate the impact of baseline calibration, which is used in quantitative cardiac MRI perfusion analysis to correct for surface coil inhomogeneity and noise, on myocardial perfusion reserve index (MPRI) and its contribution to previously reported paradoxical low MPRI < 1.0 in patients with unobstructed coronary arteries. Semiquantitative perfusion analysis was performed in 20 patients with unobstructed coronary arteries undergoing stress/rest perfusion CMR and in ten patients undergoing paired rest perfusion CMR. The following baseline calibration settings were compared: (1) baseline division, (2) baseline subtraction and (3) no baseline calibration. In uncalibrated analysis, we observed ~ 20% segmental dispersion of signal intensity (SI)-over-time curves. Both baseline subtraction and baseline division reduced relative dispersion of t0-SI (p < 0.001), but only baseline division corrected for dispersion of peak-SI and maximum upslope also (p < 0.001). In the assessment of perfusion indices, however, baseline division resulted in paradoxical low MPRI (1.01 ± 0.23 vs. 1.63 ± 0.38, p < 0.001) and rest perfusion index (RPI 0.54 ± 0.07 vs. 0.94 ± 0.12, p < 0.001), respectively. This was due to a reversed ratio of blood-pool and myocardial baseline-SI before the second perfusion study caused by circulating contrast agent from the first injection. In conclusion, baseline division reliably corrects for inhomogeneity of the surface coil sensitivity profile facilitating comparisons of regional myocardial perfusion during hyperemia or at rest. However, in the assessment of MPRI, baseline division can lead to paradoxical low results (even MPRI < 1.0 in patients with unobstructed coronary arteries) potentially mimicking severely impaired perfusion reserve. Thus, in the assessment of MPRI we propose to waive baseline calibration.

Indexed as

Magnetic Resonance Imaging, CineAdenosineAdultAgedAngina PectorisCalibrationCoronary CirculationFemaleHumansMaleMiddle AgedMyocardial Perfusion ImagingMyocarditisPredictive Value of TestsProspective StudiesReproducibility of ResultsAdenosineVasodilator AgentsAdenosine stress perfusion CMRBaseline calibrationBaseline correctionMPRIMyocardial perfusion reserveSurface coil sensitivity profile

Identifiers

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked