Repeatability and reproducibility of prospective motion- and shim corrected 2D glycoCEST MRI.
Article in Quantitative imaging in medicine and surgery, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
7 citing papers in PubMed.
- Dynamic glucose-enhanced MRI in brain tumors.Magma (New York, N.Y.) · 2026Review
- Recent developments in translational imaging of in vivo gene therapy outcomes.Molecular therapy : the journal of the American Society of Gene Therapy · 2025Review
- Amide proton transfer weighted MRI measurements yield consistent and repeatable results in patients with gliomas: a prospective test-retest study.European radiology · 2025Article
- Motion and magnetic field inhomogeneity correction techniques for chemical exchange saturation transfer (CEST) MRI: A contemporary review.NMR in biomedicine · 2025Review
- In vivo imaging of glycogen in human muscle.Nature communications · 2024Article
- Review and consensus recommendations on clinical APT-weighted imaging approaches at 3T: Application to brain tumors.Magnetic resonance in medicine · 2022Review
- Whole-brain amide CEST imaging at 3T with a steady-state radial MRI acquisition.Magnetic resonance in medicine · 2021Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
Abstract
backgroundRepeated glycoCEST MRI measurements on the same subject should produce similar results under the same environmental and experimental conditions. However, fluctuations in the static B
methodsIn nine subjects, double volumetric navigated (DvNav)-CEST acquisitions in the calf muscle were repeated five times in each of two sessions-the first without correction, and the second with real-time shim- and motion correction applied. In both sessions a dynamically changing field was introduced by running a 5-minute gradient intensive diffusion sequence. We evaluated the effect of the introduced B
resultsWith real-time shim- and motion correction, glycoCEST results were relatively consistent with mean coefficient of variation (CoV) 2.7%±1.4% across all subjects, whereas without correction the results were less consistent with CoV 84%±71%.
conclusionsOur results demonstrate that real-time shim- and motion correction can mitigate effects of B0 field fluctuations and improve reproducibility of glycoCEST data. This is important when conducting longitudinal studies or when using glycoCEST MRI to assess treatment or physiological responses over time.