Validation of differences in backscatter coefficients among four ultrasound scanners with different beamforming methods.

Masaaki Omura, Hideyuki Hasegawa, Ryo Nagaoka, Kenji Yoshida, Tadashi Yamaguchi

Abstract readValidation Study
PubMed Publisher
In one paragraph

Article in Journal of medical ultrasonics (2001), 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
1.0field-weighted citation impact, top 22% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 15 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 2 institutions in 1 country.

Masaaki OmuraGraduate School of Science and Engineering (Frontier Science Program), Chiba University, 1-33 Yayoicho, Inage, Chiba, 2638522, Japan. m.omura@chiba-u.jp.ORCID http://orcid.org/0000-0002-4432-8548
Hideyuki HasegawaGraduate School of Science and Engineering, University of Toyama, Gofuku 3190, Toyama, 9308555, Japan.
Ryo NagaokaGraduate School of Science and Engineering, University of Toyama, Gofuku 3190, Toyama, 9308555, Japan.
Kenji YoshidaCenter for Frontier Medical Engineering, Chiba University, 1-33 Yayoicho, Inage, Chiba, 2638522, Japan.
Tadashi YamaguchiCenter for Frontier Medical Engineering, Chiba University, 1-33 Yayoicho, Inage, Chiba, 2638522, Japan. yamaguchi@faculty.chiba-u.jp.
Chiba University · JPUniversity of Toyama · JP

Funding

Chiba University (JP) EngineeringChiba University (JP) Financial Support from the Frontier Science Program of Graduate School of ScienceChiba University (JP) Financial Support from the Institute for Global Prominent ResearchJapan Society for the Promotion of Science JSPS Core-to-Core Program (A. Advanced Research Networks)Japan Society for the Promotion of Science KAKENHI Grant Number 17H05280Japan Society for the Promotion of Science KAKENHI Grant Number 17J07762
6 · The paper itself

Abstract

purposeThe backscatter coefficient (BSC) indicates the absolute scatterer property of a material, independently of clinicians and system settings. Our study verified that the BSC differed among the scanners, transducers, and beamforming methods used for quantitative ultrasound analyses of biological tissues.

methodsMeasurements were performed on four tissue-mimicking homogeneous phantoms containing spherical scatterers with mean diameters of 20 and 30 µm prepared at concentrations of 0.5 and 2.0 wt%, respectively. The BSCs in the different systems were compared using ultrasound scanners with two single-element transducers and five linear high- or low-frequency probes. The beamforming methods were line-by-line formation using focused imaging (FI) and parallel beam formation using plane wave imaging (PWI). The BSC of each system was calculated by the reference phantom method. The mean deviation from the theoretical BSC computed by the Faran model was analyzed as the benchmark validation of the calculated BSC.

resultsThe BSCs calculated in systems with different properties and beamforming methods well concurred with the theoretical BSC. The mean deviation was below ± 2.8 dB on average, and within the approximate standard deviation (± 2.2 dB at most) in all cases. These variations agreed with a previous study in which the largest error among four different scanners with FI beamforming was 3.5 dB.

conclusionThe BSC in PWI was equivalent to those in the other systems and to those of FI beamforming. This result indicates the possibility of ultra-high frame-rate BSC analysis using PWI.

Indexed as

AlgorithmsPhantoms, ImagingTransducersUltrasonographyBackscatter coefficientFocused imagingLinear phased array transducerPlane wave imagingSingle-element transducer

Identifiers

PMID31679096
OpenAlexW2987941443

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.