Evidence map›Paper›PMID 31667027›Full record

ArticleIEEE journal of translational engineering in health and medicine2019

A Miniature Dual-Biomarker-Based Sensing and Conditioning Device for Closed-Loop DBS.

Mahboubeh Parastarfeizabadi, Abbas Z Kouzani

Open access · goldAbstract read
In one paragraph

Article in IEEE journal of translational engineering in health and medicine, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
0.6field-weighted citation impact, top 30% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 6 citations in OpenAlex.

  1. Closing the loop in DBS: A data-driven approach.Parkinsonism & related disorders · 2025
    Review
  2. Review
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors at 1 institution in 1 country.

Mahboubeh ParastarfeizabadiSchool of EngineeringDeakin UniversityGeelongVIC3216Australia.
Abbas Z KouzaniSchool of EngineeringDeakin UniversityGeelongVIC3216Australia.
Deakin University · AU

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

In this paper, a dual-biomarker-based neural sensing and conditioning device is proposed for closing the feedback loop in deep brain stimulation devices. The device explores both local field potentials (LFPs) and action potentials (APs) as measured biomarkers. It includes two channels, each having four main parts: (1) a pre-amplifier with built-in low-pass filter, (2) a ground shifting circuit, (3) an amplifier with low-pass function, and (4) a high-pass filter. The design specifications include miniature-size, light-weight, and 100 dB gain in the LFP and AP channels. This device has been validated through bench and in-vitro tests. The bench tests have been performed using different sinusoidal signals and pre-recorded neural signals. The in-vitro tests have been conducted in the saline solution that mimics the brain environment. The total weight of the device including a 3 V coin battery, and battery holder is 1.2 g. The diameter of the device is 11.2 mm. The device can be used to concurrently sense LFPs and APs for closing the feedback loop in closed-loop deep brain stimulation systems. It provides a tetherless head-mountable platform suitable for pre-clinical trials.

Indexed as

Analog circuitbrain sensordeep brain stimulationfabricationmultiple biomarkers

Identifiers

PMID31667027
PMCPMC6752632
OpenAlexW2970897407

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.