Evidence map›Paper›PMID 31649984›Full record

ArticlePractical laboratory medicine2019

Establishment of a stable sampling method to assay mercaptoalbumin/non-mercaptoalbumin and reference ranges.

Keiko Yasukawa, Masaya Sato, Takahiro Nojiri, Naoyuki Yoshikawa, Kazuharu Morita, Miyoko Kai, Yumiko Oike, Tsutomu Yamazaki, Hitoshi Ikeda, Yutaka Yatomi

Open access · goldAbstract read
In one paragraph

Article in Practical laboratory medicine, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.

0numbers the graph read from it
0cells of the map it votes in
7citing papers in PubMed
1.0field-weighted citation impact, top 25% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

7 citing papers in PubMed, 11 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Review
  7. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

10 authors at 1 institution in 1 country.

Keiko YasukawaDepartment of Clinical Laboratory, The University of Tokyo Hospital, Tokyo, Japan.
Masaya SatoDepartment of Clinical Laboratory, The University of Tokyo Hospital, Tokyo, Japan.
Takahiro NojiriDepartment of Clinical Laboratory, The University of Tokyo Hospital, Tokyo, Japan.
Naoyuki YoshikawaDepartment of Clinical Laboratory, The University of Tokyo Hospital, Tokyo, Japan.
Kazuharu MoritaDepartment of Clinical Laboratory, The University of Tokyo Hospital, Tokyo, Japan.
Miyoko KaiDepartment of Center for Epidemiology and Preventive, The University of Tokyo Hospital, Tokyo, Japan.
Yumiko OikeDepartment of Center for Epidemiology and Preventive, The University of Tokyo Hospital, Tokyo, Japan.
Tsutomu YamazakiDepartment of Center for Epidemiology and Preventive, The University of Tokyo Hospital, Tokyo, Japan.
Hitoshi IkedaDepartment of Clinical Laboratory, The University of Tokyo Hospital, Tokyo, Japan.
Yutaka YatomiDepartment of Clinical Laboratory, The University of Tokyo Hospital, Tokyo, Japan.
University of Tokyo Hospital · JP

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Oxidative stress is reportedly associated with many diseases such as cancer, arteriosclerosis, diabetes and aging, but no practical biomarkers are currently available in actual clinical practice. Human mercaptoalbumin (HMA) and human non-mercaptoalbumin (HNA) are expected to become markers of oxidative stress, but the stability of HMA/HNA has been problematic. We investigated the conditions for stabilizing HMA/HNA and found that HMA/HNA was stable at room temperature for 25 h if whole blood samples were mixed with a citrate buffer so that the citric acid concentration after mixing was 70 mM or higher and the pH of the added buffer was less than pH 6.0. Whole blood samples were then collected under the above conditions, and the reference range for HNA was set at 21.8% ± 7.4% (HMA, 78.2% ± 7.4%) based on samples from 65 volunteers (28 males and 37 females; average age, 55.0 ± 13.8 years). The clinical usefulness of HMA/HNA as an oxidative stress marker should be clarified for specific pathological conditions using the previously reported, highly accurate measurement method under the conditions required for HMA/HNA stability.

Indexed as

AlbuminHMA/HNAOxidative stressReference rangeStabilization method

Identifiers

PMID31649984
PMCPMC6804495
OpenAlexW2973128240

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked