Evidence map›Paper›PMID 31569017›Full record

ReviewBiomaterials2019

Mechanical considerations for polymeric heart valve development: Biomechanics, materials, design and manufacturing.

Richard L Li, Jonathan Russ, Costas Paschalides, Giovanni Ferrari, Haim Waisman, Jeffrey W Kysar, David Kalfa

Abstract readReview
In one paragraph

Review in Biomaterials, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 40 papers.

0numbers the graph read from it
0cells of the map it votes in
40citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

40 citing papers in PubMed.

  1. Review
  2. Review
  3. Article
  4. Article
  5. Article
  6. [Numerical simulation study on the influence of free edge configuration on the performance of polymeric heart valves].Sheng wu yi xue gong cheng xue za zhi = Journal of biomedical engineering = Shengwu yixue gongchengxue zazhi · 2026
    Article
  7. Material Safety of Styrene-Block-Ethylene/Butylene-Block-Styrene Copolymers Used for Cardiac Valves: 6-Month In Vivo Results from a Juvenile Sheep Model.European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery · 2025
    Article
  8. Article
  9. Review
  10. Article
  11. Review
  12. Article
  13. Review
  14. Review
  15. Review
  16. Article
  17. Article
  18. Article
  19. Article
  20. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Richard L LiDepartment of Mechanical Engineering, Fu Foundation School of Engineering and Applied Science, Columbia University, New York, NY, USA; Division of Cardiac, Thoracic and Vascular Surgery, Section of Pediatric and Congenital Cardiac Surgery, New-York Presbyterian - Morgan Stanley Children's Hospital, Columbia University Medical Center, New York, NY, USA.
Jonathan RussDepartment of Civil Engineering and Engineering Mechanics, Fu Foundation School of Engineering and Applied Science, Columbia University, New York, NY, USA.
Costas PaschalidesDepartment of Mechanical Engineering, Fu Foundation School of Engineering and Applied Science, Columbia University, New York, NY, USA.
Giovanni FerrariDepartment of Surgery and Biomedical Engineering, Columbia University Medical Center, New York, NY, USA.
Haim WaismanDepartment of Civil Engineering and Engineering Mechanics, Fu Foundation School of Engineering and Applied Science, Columbia University, New York, NY, USA.
Jeffrey W KysarDepartment of Mechanical Engineering, Fu Foundation School of Engineering and Applied Science, Columbia University, New York, NY, USA; Department of Otolaryngology - Head and Neck Surgery, Columbia University Medical Center, New York, NY, USA. Electronic address: jk2079@columbia.edu.
David KalfaDivision of Cardiac, Thoracic and Vascular Surgery, Section of Pediatric and Congenital Cardiac Surgery, New-York Presbyterian - Morgan Stanley Children's Hospital, Columbia University Medical Center, New York, NY, USA. Electronic address: dk2757@cumc.columbia.edu.

Funding

Oxidation-mediated structural degeneration of bioprosthetic heart valvesR01HL143008 · NHLBI · CHILDREN'S HOSP OF PHILADELPHIA · PI FERRARI, GIOVANNI, LEVY, ROBERT J · 2018 to 2021
$3.2M
NHLBI NIH HHS R01 HL143008
6 · The paper itself

Abstract

The native human heart valve leaflet contains a layered microstructure comprising a hierarchical arrangement of collagen, elastin, proteoglycans and various cell types. Here, we review the various experimental methods that have been employed to probe this intricate microstructure and which attempt to elucidate the mechanisms that govern the leaflet's mechanical properties. These methods include uniaxial, biaxial, and flexural tests, coupled with microstructural characterization techniques such as small angle X-ray scattering (SAXS), small angle light scattering (SALS), and polarized light microscopy. These experiments have revealed complex elastic and viscoelastic mechanisms that are highly directional and dependent upon loading conditions and biochemistry. Of all engineering materials, polymers and polymer-based composites are best able to mimic the tissue-level mechanical behavior of the native leaflet. This similarity to native tissue permits the fabrication of polymeric valves with physiological flow patterns, reducing the risk of thrombosis compared to mechanical valves and in some cases surpassing the in vivo durability of bioprosthetic valves. Earlier work on polymeric valves simply assumed the mechanical properties of the polymer material to be linear elastic, while more recent studies have considered the full hyperelastic stress-strain response. These material models have been incorporated into computational models for the optimization of valve geometry, with the goal of minimizing internal stresses and improving durability. The latter portion of this review recounts these developments in polymeric heart valves, with a focus on mechanical testing of polymers, valve geometry, and manufacturing methods.

Indexed as

Heart Valve ProsthesisProsthesis DesignAnimalsBiomechanical PhenomenaHeart ValvesHumansPolymersStress, MechanicalPolymersMechanical testingMechanobiologyNative heart valvesPolymeric heart valvesSoft tissue mechanics

Identifiers

PMID31569017
PMCPMC6948849

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.