A comparison of IROC and ACDS on-site audits of reference and non-reference dosimetry.
Article in Medical physics, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
7 citing papers in PubMed, 17 citations in OpenAlex.
- Multicentre dosimetry audit in the European randomized phase III study of neoadjuvant chemoradiotherapy using proton versus photon radiotherapy in locally advanced oesophageal cancer.Physics and imaging in radiation oncology · 2026Article
- Results of a remote audit program of reference and non-reference dosimetry of photon beams using alanine detectors.Physics and imaging in radiation oncology · 2026Article
- Developing reference plans for evaluating global clinical trials credentialing and PSQA systems.Journal of applied clinical medical physics · 2025Article
- In Australia professional registration for qualified medical physicists should be mandated through the Australian Health Practitioner Regulation Agency (AHPRA).Physical and engineering sciences in medicine · 2024Article
- Optimised conformal total body irradiation: a heterogeneous practice, so where next?The British journal of radiology · 2023Review
- Adoption of respiratory motion management in radiation therapy.Physics and imaging in radiation oncology · 2022Article
- Commissioning measurements on an Elekta Unity MR-Linac.Physical and engineering sciences in medicine · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors at 3 institutions in 2 countries.
Funding
Abstract
purposeConsistency between different international quality assurance groups is important in the progress toward similar standards and expectations in radiotherapy dosimetry around the world, and in the context of consistent clinical trial data from international trial participants. This study compares the dosimetry audit methodology and results of two international quality assurance groups performing a side-by-side comparison at the same radiotherapy department, and interrogates the ability of the audits to detect deliberately introduced errors.
methodsA comparison of the core dosimetry components of reference and non-reference audits was conducted by the Imaging and Radiation Oncology Core (IROC, Houston, USA) and the Australian Clinical Dosimetry Service (ACDS, Melbourne, Australia). A set of measurements were conducted over 2 days at an Australian radiation therapy facility in Melbourne. Each group evaluated the reference dosimetry, output factors, small field output factors, percentage depth dose (PDD), wedge, and off-axis factors according to their standard protocols. IROC additionally investigated the Electron PDD and the ACDS investigated the effect of heterogeneities. In order to evaluate and compare the performance of these audits under suboptimal conditions, artificial errors in percentage depth dose (PDD), EDW, and small field output factors were introduced into the 6 MV beam model to simulate potential commissioning/modeling errors and both audits were tested for their sensitivity in detecting these errors.
resultsWith the plans from the clinical beam model, almost all results were within tolerance and at an optimal pass level. Good consistency was found between the two audits as almost all findings were consistent between them. Only two results were different between the results of IROC and the ACDS. The measurements of reference FFF photons showed a discrepancy of 0.7% between ACDS and IROC due to the inclusion of a 0.5% nonuniformity correction by the ACDS. The second difference between IROC and the ACDS was seen with the lung phantom. The asymmetric field behind lung measured by the ACDS was slightly (0.3%) above the ACDS's pass (optimal) level of 3.3%. IROC did not detect this issue because their measurements were all assessed in a homogeneous phantom. When errors were deliberately introduced neither audit was sensitive enough to pick up a 2% change to the small field output factors. The introduced PDD change was flagged by both audits. Similarly, the introduced error of using 25° wedge instead of 30° wedge was detectible in both audits as out of tolerance.
conclusionsDespite different equipment, approach, and scope of measurements in on-site audits, there were clear similarities between the results from the two groups. This finding is encouraging in the context of a global harmonized approach to radiotherapy quality assurance and dosimetry audit.