Accuracy of Glucose Meter among Adults in a Semi-urban Area in Kathmandu, Nepal.
Article in JNMA; journal of the Nepal Medical Association. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
4 citing papers in PubMed, 4 citations in OpenAlex.
- Assessment of analytical performance of a glucometer in Indian adults at a tertiary care hospital.BMC endocrine disorders · 2025Article
- Using easy-to-collect indices to develop and validate models for identifying metabolic syndrome and pre-metabolic syndrome.Frontiers in endocrinology · 2025Article
- Accuracy of FreeStyle Libre flash glucose monitoring in patients with type 2 diabetes who migrated from highlands to plains.World journal of diabetes · 2024Article
- Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
introductionGlucose meters are gaining popularity in monitoring of blood glucose at household levels and in health care set-ups due to their portability, affordability and convenience of use over the laboratory based reference methods. Still they are not free of limitations. Operator's technique, extreme temperatures, humidity, patients' medication, hematocrit values can affect the reliability of glucose meter results. Hence, the accuracy of glucose meter has been the topic of concern since years. Therefore, present study aims to evaluate the analytical and clinical accuracy of glucose meter using International Organization for Standardization 15197 guideline.
methodsA community based descriptive cross-sectional study was conducted in Kapan, Kathmandu, Nepal in April 2018. Glucose levels were measured using glucose meter and reference laboratory method simultaneously among 203 adults ≥20 years, after an overnight fasting and two hours of ingestion of 75 grams glucose. Modified Bland-Altman plots were created by incorporating ISO 15197 guidelines to check the analytical accuracy and Park error grid was used to evaluate the clinical accuracy of the device.
resultsModified Bland-Altman plots showed>95% of the test results were beyond the acceptable analytical criteria of ISO 15197:2003 and 2013. Park Error Grid-Analysis showed 99% of the data within zones A and B of the consensus error grid.
conclusionsGlucose meter readings were within clinically acceptable parameters despite discrepancies on analytical merit. Possible sources of interferences must be avoided during the measurement to minimize the disparities and the values should be interpreted with caution.