Trial reportInternational urology and nephrology2019
Reversed connection of cuffed, tunneled, dual-lumen catheters with increased blood flow rate maintains the adequacy of delivered dialysis despite the higher access recirculation.
Trial report in International urology and nephrology, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed, 7 citations in OpenAlex.
- Reliability of monitoring acid-base and electrolyte parameters through circuit lines during regional citrate anticoagulation-continuous renal replacement therapy.Nursing in critical care · 2022Article
- Effect of increased blood flow rate on renal anemia and hepcidin concentration in hemodialysis patients.BMC nephrology · 2021Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeReversing the connection mode of permanent dual-lumen dialysis catheters results in higher access recirculation that may compromise the dialysis adequacy. The purpose of this study is to investigate the effect of reversed-connected dialysis with a higher versus a standard blood flow rate (Qb) on adequacy parameters and access recirculation. MATERIALS AND
methodsIn a cross-over design, 46 prevalent dialysis patients with a properly functioning cuffed, tunneled, dual-lumen catheter were evaluated in three consecutive mid-week dialysis sessions. At baseline, participants were evaluated under standardized conditions (correct connection, Qb = 300 ml/min). In Phase A, dialysis was performed with reversed connection and Qb = 300 ml/min. In Phase B, dialysis was performed with reversed connection and Qb = 400 ml/min. The sequence of evaluations (Phase A and B or vise verse) was randomized. All other dialysis-related parameters were unchanged in all three occasions.
resultsAs expected, compared with baseline, reversed-connected dialysis in Phase A resulted in lower URR and spKt/V, and in a higher recirculation rate. Compared with baseline, reversed-connected dialysis with a higher Qb in Phase B resulted in an even higher recirculation rate, but the parameters of dialysis adequacy were not different. Increase in Qb from 300 to 400 ml/min resulted in an improvement of the dialysis adequacy (URR: 64.1 ± 7.8% vs. 70.6 ± 8.2%, P < 0.001; spKt/V: 1.22 ± 0.3 vs. 1.45 ± 0.3, P < 0.001) despite the higher recirculation rate.
conclusionThis study suggests that reversed-connected dialysis with increasing Qb maintains the adequacy of the delivered dialysis despite the compensatory increase in recirculation.