Evidence map›Paper›PMID 31440282›Full record

ArticleARYA atherosclerosis2019

The efficacy and safety of using amplatzer for transcatheter closure of atrial septal defect in small children with less than 10 kg.

Mehdi Ghaderian, Mohammad Reza Sabri, Ali Reza Ahmadi, Bahar Dehghan, Chehre Mahdavi, Zakie Zahra Ataei

Abstract read
In one paragraph

Article in ARYA atherosclerosis, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed.

  1. Review
  2. Article
  3. Review
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Mehdi GhaderianAssociate Professor, Pediatric Cardiovascular Research Center, Cardiovascular Research Institute, Isfahan University of Medical Sciences, Isfahan, Iran.
Mohammad Reza SabriProfessor, Pediatric Cardiovascular Research Center, Cardiovascular Research Institute, Isfahan University of Medical Sciences, Isfahan, Iran.
Ali Reza AhmadiAssociate Professor, Pediatric Cardiovascular Research Center, Cardiovascular Research Institute, Isfahan University of Medical Sciences, Isfahan, Iran.
Bahar DehghanAssistant Professor, Pediatric Cardiovascular Research Center, Cardiovascular Research Institute, Isfahan University of Medical Sciences, Isfahan, Iran.
Chehre MahdaviAssistant Professor, Pediatric Cardiovascular Research Center, Cardiovascular Research Institute, Isfahan University of Medical Sciences, Isfahan, Iran.
Zakie Zahra AtaeiStudent of Medicine, Cardiac Rehabilitation Research Center, Cardiovascular Research Institute, Isfahan University of Medical Sciences, Isfahan, Iran.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundAtrial septal defect (ASD) accounts for about 10% of congenital heart diseases (CHDs). Self-closure of these defects in patients with defects less than 8 mm has been reported in several studies. In children, transcatheter closure of the ASD is suggested for asymptomatic patients older than two years and with weight > 15 kg. The purpose of this study was to show that transcatheter closure of ASD in small children with body weight less than 10 kg is an effective and safe method.

methodsBetween July 2016 and September 2018, 35 children with body weight less than 10 kg underwent percutaneous closure of ASD using amplatzer. All patients had minimum defect size of 6 mm, pulmonary blood flow (Qp) to systemic blood flow (Qs) ratio above 1.5, right atrial and ventricular dilation, symptoms of delayed growth, and recurrent respiratory infections in their evaluation and had acceptable rims for intervention. Follow-up evaluations were done 1 day, 1 week, 1 month, 6 months, and yearly after discharge with transthoracic echocardiography (TTE) and electrocardiography (ECG).

resultsThe mean age of patients at procedure was 12.06 ± 4.47 months (range: 6 to 14 months), mean weight was 8.32 ± 0.72 kg (range: 7.5 to 9.8 kg). The mean defect size was 10.00 ± 2.32 mm (range: 6-13 mm). The mean device size used was 10.57 ± 2.57 mm (range: 7.5 to 15 mm). Mean duration of follow-up was 16.66 ± 6.93 months (range: 1-29 months). Respiratory rate, heart rate, pulmonary stenosis (PS), and Qp to Qs ratio had significant difference before and after procedure during the follow up (P < 0.001).

conclusionTranscatheter closure of ASD with amplatzer in symptomatic small children and infants is a safe and effective treatment associated with excellent success, but long-term follow-up in a large number of patients would be warranted.

Indexed as

Atrial Septal DefectDevicesSeptal Occluder

Identifiers

PMID31440282
PMCPMC6597800

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.