Evidence map›Paper›PMID 31379983›Full record

ArticleProceedings of IEEE Sensors. IEEE International Conference on Sensors2018

Towards Bimanual Robot-Assisted Retinal Surgery: Tool-to-Sclera Force Evaluation.

Changyan He, Marina Roizenblatt, Niravkumar Patel, Ali Ebrahimi, Yang Yang, Peter L Gehlbach, Iulian Iordachita

Abstract read
In one paragraph

Article in Proceedings of IEEE Sensors. IEEE International Conference on Sensors, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 10 papers.

0numbers the graph read from it
0cells of the map it votes in
10citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

10 citing papers in PubMed.

  1. Exploring the Needle Tip Interaction Force with Retinal Tissue Deformation in Vitreoretinal Surgery.IEEE International Conference on Robotics and Automation : ICRA : [proceedings]. IEEE International Conference on Robotics and Automation · 2024
    Article
  2. Force-based Control for Safe Robot-assisted Retinal Interventions:IEEE transactions on medical robotics and bionics · 2022
    Article
  3. Robotic Assistance for Intraocular Microsurgery: Challenges and Perspectives.Proceedings of the IEEE. Institute of Electrical and Electronics Engineers · 2022
    Article
  4. Scleral Force Evaluation During Vitreoretinal Surgery: in an In Vivo Rabbit Eye Model.Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference · 2020
    Article
  5. Toward Improving Patient Safety and Surgeon Comfort in a Synergic Robot-Assisted Eye Surgery: A Comparative Study.Proceedings of the ... IEEE/RSJ International Conference on Intelligent Robots and Systems. IEEE/RSJ International Conference on Intelligent Robots and Systems · 2019
    Article
  6. Towards securing the sclera against patient involuntary head movement in robotic retinal surgery.RO-MAN ... : the ... IEEE International Symposium on Robot and Human Interactive Communication : proceedings. IEEE International Symposium on Robot and Human Interactive Communication · 2019
    Article
  7. Sclera Force Evaluation During Vitreoretinal Surgeries in Ex Vivo Porcine Eye Model.Proceedings of IEEE Sensors. IEEE International Conference on Sensors · 2019
    Article
  8. Dual-Stiffness Force-Sensing Cannulation Tool for Retinal Microsurgery.Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference · 2019
    Article
  9. Adaptive Control of Sclera Force and Insertion Depth for Safe Robot-Assisted Retinal Surgery.IEEE International Conference on Robotics and Automation : ICRA : [proceedings]. IEEE International Conference on Robotics and Automation · 2019
    Article
  10. Sclera Force Control in Robot-assisted Eye Surgery: Adaptive Force Control vs. Auditory Feedback.... International Symposium on Medical Robotics. International Symposium on Medical Robotics · 2019
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Changyan HeLCSR at the Johns Hopkins University, Baltimore, MD 21218 USA.
Marina RoizenblattWilmer Eye Institute at the Johns Hopkins Hospital, Baltimore, MD 21287 USA.
Niravkumar PatelLCSR at the Johns Hopkins University, Baltimore, MD 21218 USA.
Ali EbrahimiLCSR at the Johns Hopkins University, Baltimore, MD 21218 USA.
Yang YangSchool of Mechanical Engineering and Automation at Beihang University, Beijing, 100191 China.
Peter L GehlbachWilmer Eye Institute at the Johns Hopkins Hospital, Baltimore, MD 21287 USA.
Iulian IordachitaLCSR at the Johns Hopkins University, Baltimore, MD 21218 USA.

Funding

An Active Handheld MicromanipulatorR01EB000526 · NIBIB · CARNEGIE-MELLON UNIVERSITY · PI RIVIERE, CAMERON N · 2003 to 2017
$2.6M
Enabling Technology for Safe Robot-assisted Surgical MicromanipulationR01EB023943 · NIBIB · JOHNS HOPKINS UNIVERSITY · PI GEHLBACH, PETER LOUIS, IORDACHITA, IULIAN IOAN · 2017 to 2024
$2.3M
NIBIB NIH HHS R01 EB000526NIBIB NIH HHS R01 EB023943
6 · The paper itself

Abstract

The performance of retinal microsurgery often requires the coordinated use of both hands. During bimanual retinal surgery, dominant hand performance may be negatively impacted by poor non-dominant hand assistance. Therefore understanding bimanual latent determinants, and establishing safety criteria for bimanual manipulation is relevant to robotic development and to eventual patient care. In this paper, we present a preliminary study to quantitatively evaluate one aspect of bimanual tool use in retinal surgery. Two force sensing tools were designed and fabricated using fiber Bragg grating sensors. Tool-to-sclera contact force is measured using the developed tools and analyzed. The tool forces were recorded during five basic surgical maneuvers typical of retinal surgery. Two subjects are involved in experiments, including one clinician and one engineer. For comparison, all manipulations were replicated under robot-assisted conditions. The results indicate that the average tool-to-sclera force recorded from the dominant hand tool is significantly higher than that from the non-dominant hand tool (

Indexed as

bimanual manipulationrobot-assisted retinal surgerytool-to-sclera force

Identifiers

PMID31379983
PMCPMC6677262

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.