Evidence map›Paper›PMID 31322701›Full record

ArticleEuropace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology2019

Clinical safety of the ProMRI implantable cardioverter-defibrillator systems during head and lower lumbar magnetic resonance imaging at 3 T: results of the ProMRI 3T ENHANCED Master study.

Rainer Zbinden, Christian Wollmann, Johannes Brachmann, Jochen Michaelsen, Clemens Steinwender, Pramesh Kovoor, Sebastian Kelle, Andrew D McGavigan, Chi Keong Ching, Gemma A Figtree and 3 more

Open access · bronzeAbstract readMulticenter Study
In one paragraph

Article in Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
0.2field-weighted citation impact, top 44% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed, 2 citations in OpenAlex.

  1. 3 Tesla magnetic resonance imaging in patients with cardiac implantable electronic devices: a single centre experience.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2023
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

13 authors at 13 institutions in 5 countries.

Rainer ZbindenKlinik für Kardiologie, Stadtspital Triemli, Birmensdorferstrasse 497, Zurich, Switzerland.
Christian WollmannKlinische Abteilung für Innere Medizin 3, Universitätsklinikum St. Pölten, Dunant-Platz 1, 3100 St. Pölten, Austria.
Johannes BrachmannII. Medizinische Klinik Kardiologie, Angiologie und Pneumologie, Klinikum Coburg, Ketschendorfer Str. 33, Coburg, Germany.
Jochen MichaelsenKlinik für Kardiologie, Angiologie und Schlafmedizin, Bonifatius Hospital Lingen GmbH, Wilhelmstr. 13, Lingen, Germany.
Clemens SteinwenderKlinik für Kardiologie, Kepler Universitätsklinikum, Krankenhausstr. 9, Linz, Austria.
Pramesh KovoorDepartment of Cardiology, Westmead Hospital, Cnr Hawkesbury Road and Darcy Road, Westmead (NSW), Australia.
Sebastian KelleDepartment of Internal Medicine/Cardiology, German Heart Center Berlin and Charité University Medicine Berlin, Augustenburger Platz 1, Berlin, Germany.
Andrew D McGaviganDepartment of Cardiology, Flinders Medical Centre, 1 Flinders Drive, Bedford Park, Australia.
Chi Keong ChingDepartment of Electrophysiology, National Heart Centre of Singapore, 5 Hospital Drive Singapore, Singapore, Singapore.
Gemma A FigtreeKolling Institute, University of Sydney, Royal North Shore Hospital, Pacific Hwy, St Leonards (NSW), Australia.
Jan SchmidtKlinik für Kardiologie, Pneumologie & Angiologie, Universitätsklinikum Düsseldorf, Moorenstr. 5, Düsseldorf, Germany.
Tobias TimmelCenter for Clinical Research, Biotronik SE & Co. KG, Woermannkehre 1, Berlin, Germany.
Joachim LotzInstituts für Diagnostische und Interventionelle Radiologie, Georg-August-Universität, Robert-Koch-Str. 40, Göttingen, Germany.
Biotronik (Germany) · DEDeutsches Herzzentrum der Charité · DEDüsseldorf University Hospital · DEFlinders Medical Centre · AUKepler Universitätsklinikum · ATKlinikum Coburg · DENational Heart Centre Singapore · SGSt. Bonifatius Hospital · DEThe University of Sydney · AUTriemli Hospital · CHUniversitätsklinikum St. Pölten · ATUniversity of Göttingen · DEWestmead Hospital · AU

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

aimsThere have been no published studies on the safety of magnetic resonance imaging (MRI) at 3 Tesla (3 T) in patients with MRI-conditional implantable cardioverter-defibrillators (ICDs). The aim of this study was to assess clinical safety of the Biotronik ProMRI ICD system during non-diagnostic head and lower lumbar scans under 3 T MRI conditions. METHODS AND

resultsThe study enrolled 129 patients at 12 sites in Australia, Singapore, and Europe. Predefined head and lower lumbar MR scans (total duration ≈30 min) were performed in 112 patients. Three primary endpoints were evaluated from the pre-MRI to the 1-month post-MRI visit: (i) freedom from serious adverse device effects (SADEs) related to MRI (hypothesized to be >90%); (ii) pacing threshold invariance for all leads (geometric mean of the patient-wise ratios for 1 month vs. pre-MRI was hypothesized to be <1.07); and (iii) sensing amplitude invariance (geometric mean of the ratios was hypothesized to be >0.993). No MRI-related SADE occurred (SADE-free rate 100%, 95% confidence interval 95.98-100%). Pacing threshold and sensing amplitudes fulfilled the invariance hypotheses with high statistical significance (P < 0.0013). No threshold increase >0.5 V or sensing amplitude decrease by >50% was observed (secondary endpoints). Lead impedances, battery capacity, and detection and treatment of arrhythmias by ICDs were not affected by MRI scans.

conclusionThe head and lower lumbar scans under specific 3 T MRI conditions were safe in the investigated MR-conditional ICD systems. There was no evidence of harm to the patients or any negative influence of the MRI scan on the implanted systems.

Indexed as

Defibrillators, ImplantableAdultAgedAged, 80 and overArrhythmias, CardiacEquipment DesignEquipment SafetyFemaleFollow-Up StudiesHeadHeart RateHumansLumbar VertebraeMagnetic Resonance ImagingMaleMiddle AgedElectromagnetic interferenceImplantable cardioverter-defibrillatorMagnetic resonance imagingProMRISafety

Identifiers

PMID31322701
PMCPMC6826205
OpenAlexW2962992064

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.