ArticleJMIR research protocols2019
Design of Improved Intertrochanteric Fracture Treatment (DRIFT) Study: Protocol for Biomechanical Testing and Finite Element Analysis of Stable and Unstable Intertrochanteric Fractures Treated With Intramedullary Nailing or Dynamic Compression Screw.
Article in JMIR research protocols, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
7 citing papers in PubMed, 1 synthesis or guideline pooled it, 7 citations in OpenAlex.
- Implants for fixation of intertrochanteric femoral fracture: a systematic review and network meta-analysis of randomized controlled trials.BMC musculoskeletal disorders · 2025Pooled it
- Article
- Radiographic Predictors of Leg Shortening After Curved Intertrochanteric Varus Osteotomy in Osteonecrosis of the Femoral Head.Orthopaedic surgery · 2025Article
- Inverse relationship between femoral lateralization and neck-shaft angle is a joint event after intramedullary nailing of per trochanteric fractures.Scientific reports · 2023Article
- Standardized artificially created stable pertrochanteric femur fractures present more homogenous results compared to osteotomies for orthopaedic implant testing.BMC musculoskeletal disorders · 2021Article
- Loss of Reduction after Cephalomedullary Nail Fixation of Intertrochanteric Femoral Fracture: A Brief Report.Orthopaedic surgery · 2020Article
- Clinical and functional comparison of dynamic hip screws and intramedullary nails for treating proximal femur metastases in older individuals.Chinese journal of cancer research = Chung-kuo yen cheng yen chiu · 2020Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 3 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundIntertrochanteric hip fractures rank in the top 10 of all impairments worldwide in terms of loss in disability-adjusted years for people aged older than 60 years. The type of surgery is usually carried out with dynamic hip screw (DHS) devices or cephalomedullary nails (CMN). Cut-out of the hip screw is considered the most frequent mechanical failure for all implants with an estimated incidence ranging from 2% to 16.5%; this entails both enhancing our understanding of the prognostic factors of cut-out and improving all aspects of intertrochanteric fracture treatment.
objectiveThe Design of Improved Intertrochanteric Fracture Treatment (DRIFT) study's main objective is to provide intertrochanteric fracture treatment expertise, requirements and specifications, clinical relevance, and validation to improve treatment outcomes by developing a universal algorithm for designing patient- and fracture-oriented treatment. The hypothesis to be tested is that a more valgus reduction angle and implants of higher angles will lead to a more favorable biomechanical environment for fracture healing-that is, higher compressive loads at the fracture site with lower shear loads at the hip screw femoral head interface. A new implant with enhanced biomechanical and technical characteristics will be designed and fabricated; in addition, an integrated design and optimization platform based on computer-aided design tools and topology optimization modules will be developed.
methodsTo test this hypothesis, a biomechanical study comprising experimental loading of synthetic femora (Sawbones Inc) and finite element analysis (FEA) will be conducted. Detailed FEA of existing implants (DHS and CMN) implemented in different clinical cases under walking conditions will be performed to derive the stress and strain fields developed at the implant-bone system and identify critical scenarios that could lead to failure of therapy. These models would be validated against instrumented mechanical tests using strain gages and a digital image correlation process.
resultsAfter testing, geometric drawbacks of existing implants will be fully recognized, and geometric characteristics will be correlated with critical failure scenarios. The last step would be the numeric design, computer-aided design (using FEA codes and design packages), and optimization of the new proposed implant with regard to improved biomechanical surgical technique and enhanced mechanical performance that will reduce the possibility for critical failure scenarios.
conclusionsThe optimization of the biomechanical behavior of the fracture-osteosynthesis model by the application of the ideal reduction angle and implant is expected to have a positive effect to the rate of mechanical failure and, subsequently, the healing rates, morbidity, and mortality in this fragile patient group. INTERNATIONAL REGISTERED REPORT IDENTIFIER (IRRID): DERR1-10.2196/12845.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials