Evidence map›Paper›PMID 31266252›Full record

ArticleSensors (Basel, Switzerland)2019

Simulating Arbitrary Electrode Reversals in Standard 12-lead ECG.

Vessela Krasteva, Irena Jekova, Ramun Schmid

Abstract read
In one paragraph

Article in Sensors (Basel, Switzerland), 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed.

  1. Article
  2. Smart Sensors for Healthcare and Medical Applications.Sensors (Basel, Switzerland) · 2021
    Article
  3. Article
  4. Electrocardiographic lead reversals.Indian pacing and electrophysiology journal
    Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors.

Vessela KrastevaInstitute of Biophysics and Biomedical Engineering, Bulgarian Academy of Sciences, Acad. G. Bonchev Str. Bl 105, 1113 Sofia, Bulgaria. vessika@biomed.bas.bg.
Irena JekovaInstitute of Biophysics and Biomedical Engineering, Bulgarian Academy of Sciences, Acad. G. Bonchev Str. Bl 105, 1113 Sofia, Bulgaria.
Ramun SchmidSignal Processing, Schiller AG, Altgasse 68, CH-6341 Baar, Switzerland.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Electrode reversal errors in standard 12-lead electrocardiograms (ECG) can produce significant ECG changes and, in turn, misleading diagnoses. Their detection is important but mostly limited to the design of criteria using ECG databases with simulated reversals, without Wilson's central terminal (WCT) potential change. This is, to the best of our knowledge, the first study that presents an algebraic transformation for simulation of all possible ECG cable reversals, including those with displaced WCT, where most of the leads appear with distorted morphology. The simulation model of ECG electrode swaps and the resultant WCT potential change is derived in the standard 12-lead ECG setup. The transformation formulas are theoretically compared to known limb lead reversals and experimentally proven for unknown limb-chest electrode swaps using a 12-lead ECG database from 25 healthy volunteers (recordings without electrode swaps and with 5 unicolor pairs swaps, including red (right arm-C1), yellow (left arm-C2), green (left leg (LL) -C3), black (right leg (RL)-C5), all unicolor pairs). Two applications of the transformation are shown to be feasible: 'Forward' (simulation of reordered leads from correct leads) and 'Inverse' (reconstruction of correct leads from an ECG recorded with known electrode reversals). Deficiencies are found only when the ground RL electrode is swapped as this case requires guessing the unknown RL electrode potential. We suggest assuming that potential to be equal to that of the LL electrode. The 'Forward' transformation is important for comprehensive training platforms of humans and machines to reliably recognize simulated electrode swaps using the available resources of correctly recorded ECG databases. The 'Inverse' transformation can save time and costs for repeated ECG recordings by reconstructing the correct lead set if a lead swap is detected after the end of the recording. In cases when the electrode reversal is unknown but a prior correct ECG recording of the same patient is available, the 'Inverse' transformation is tested to detect the exact swapping of the electrodes with an accuracy of (96% to 100%).

Indexed as

ElectrodesAdultAgedArmElectrocardiographyFemaleHumansMaleMiddle AgedECG electrode potentialsECG electrode swapsMSMinv transformationreconstructing correct ECG leadsunicolor limb–chest electrodesWCT potential change

Identifiers

PMID31266252
PMCPMC6651562

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.