Evidence map›Paper›PMID 31100485›Full record

ArticleJournal of the mechanical behavior of biomedical materials2019

The effect of manufacturing tolerances on the mechanical environment of taper junctions in modular TKR.

Kyle Snethen, Jorge Hernandez, Melinda Harman

Abstract read
In one paragraph

Article in Journal of the mechanical behavior of biomedical materials, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors.

Kyle SnethenBioengineering Department, Clemson University, Clemson, USA.
Jorge HernandezBioengineering Department, Clemson University, Clemson, USA.
Melinda HarmanBioengineering Department, Clemson University, Clemson, USA. Electronic address: harman2@clemson.edu.

Funding

NM-INBRE Sequencing and Bioinformatics CoreP20GM103451 · NIGMS · NEW MEXICO STATE UNIVERSITY LAS CRUCES · PI Charlotte C. Gard · 2012 to 2026
$61.2M
Tissue Structural and Neural Remodeling in Human Sacroiliac JointP20GM121342 · NIGMS · CLEMSON UNIVERSITY · PI Hai Yao · 2018 to 2026
$24.7M
NIGMS NIH HHS P20 GM103451NIGMS NIH HHS P20 GM121342
6 · The paper itself

Abstract

Taper design is known to influence corrosive behavior in taper junctions used in modular orthopaedic devices. Manufacturing tolerance of bore-cone tapers is a critical design parameter due to the effect on taper fit, but the effect of variations in manufacturing tolerance on the mechanics of taper junctions has not been well characterized, particularly in modular total knee replacement (TKR). The purpose of this study was to investigate the effect of manufacturing tolerance on stress and micromotion of modular TKR taper junctions. A 3D finite element (FE) model of a modular TKR taper junction was developed and assigned elastoplastic material properties. Model taper geometry was varied by perturbing the angle mismatch by 0.05° between ±0.25° and represented expected variation in manufacturing tolerance. Stress and micromotion were calculated during dynamic FE simulations for each taper junction geometry under varying activity loads and material combinations. Although an increase in angle mismatch generally resulted in higher stress and micromotion, plastic material behavior disrupted this trend for larger angle mismatches. Model predictions corresponded with corrosion behavior evident in vitro. If the FE results obtained here apply in vivo, the absence of elastoplastic material properties in a taper model may grossly overestimate the micromotion and underestimate corrosion behavior and ion release. It is recommended that manufacturing tolerances of bore-cone tapers in modular TKR designs should produce angle mismatches within 0.1° at the taper junction.

Indexed as

Hip ProsthesisArthroplasty, Replacement, HipArthroplasty, Replacement, KneeComputer SimulationCorrosionElasticityFinite Element AnalysisHumansImaging, Three-DimensionalMaterials TestingMotionPressureProsthesis DesignProsthesis FailureStress, MechanicalAngle mismatchFinite element analysisManufacturing toleranceMicromotionModular TKRStress