ArticleBMC pediatrics2019
First clinical experience with the Kora pacemaker system in congenital complete heart block in newborn infants.
Article in BMC pediatrics, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed, 1 citations in OpenAlex.
- Atrial auto-short phenomenon as a rare cause of ventricular lead failure in a pediatric dual chamber pacemaker patient.Pacing and clinical electrophysiology : PACE · 2020Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundTo report first clinical experience on three cases of congenital complete heart block and the use of a pacemaker system with a maximum lower rate interval of 95 beats per minute.
methodsWe retrospectively analyzed three patients treated with a pacemaker system with a maximum lower rate interval of 95 beats per minute suffering from congenital complete heart block. We report a follow up period of 2.9 years, focusing on the patients' growth, development, and adverse events, as well as pacemaker function.
resultsIn all three patients pacemaker function was impeccable, including minute ventilation sensor rate adaption. All patients showed limited growths as expected, adequate development, good feeding tolerability and circadiane heart rate adaption. One patient experienced skin traction and revision. All patients showed high aortic velocity time integral values after birth.
conclusionThe use of a pacemaker system with a maximum lower rate interval of 95 beats per minute in infants suffering from congenital complete heart block and showing high aortic VTI values seems to be feasible and to result in limited growths but adequate development.