ArticleSpine2019
Mechanical Performance of Posterior Spinal Instrumentation and Growing Rod Implants: Experimental and Computational Study.
Article in Spine, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
7 citing papers in PubMed.
- Biomechanical properties of double-headed pedicle screws and parallel connectors in thoracolumbar extension revision fixation: a finite element analysis.BMC musculoskeletal disorders · 2026Article
- Investigation of instrumentation lengthening in the thoracolumbar junction of the spine - an in vitro Biomechanical study.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2025Article
- Design and mechanical evaluation of a novel dynamic growth rod to improve the surgical treatment of early-onset scoliosis.Bone & joint research · 2025Article
- A Novel Dynamic Growth Rod Inducing Spinal Growth Modulation for the Correction of Spinal Deformities.JOR spine · 2025Article
- Simulation and validation of spinal construct testing based on ASTM F1717.Frontiers in bioengineering and biotechnology · 2025Article
- Current benchtop protocols are not appropriate for the evaluation of distraction-based growing rods: a literature review to justify a new protocol and its development.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2022Review
- Clinically relevant finite element technique based protocol to evaluate growing rods for early onset scoliosis correction.JOR spine · 2020Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
Abstract
STUDY
designExperimental and computational study of posterior spinal instrumentation and growing rod constructs per ASTM F1717-15 vertebrectomy methodology for static compressive bending.
objectiveAssess mechanical performance of standard fusion instrumentation and growing rod constructs. SUMMARY OF BACKGROUND DATA: Growing rod instrumentation utilizes fewer anchors and spans longer distances, increasing shared implant loads relative to fusion. There is a need to evaluate growing rod's mechanical performance. ASTM F1717-15 standard assesses performance of spinal instrumentation; however, effects of growing rods with side-by-side connectors have not been evaluated.
methodsStandard and growing rod constructs were tested per ASTM F1717-15 methodology; setup was modified for growing rod constructs to allow for connector offset. Three experimental groups (standard with active length 76 mm, and growing rods with active lengths 76 and 376 mm; n = 5/group) were tested; stiffness, yield load, and load at maximum displacement were calculated. Computational models were developed and used to locate stress concentrations.
resultsFor both constructs at 76 mm active length, growing rod stiffness (49 ± 0.8 N/mm) was significantly greater than standard (43 ± 0.4 N/mm); both were greater than growing rods at 376 mm (10 ± 0.3 N/mm). No significant difference in yield load was observed between growing rods (522 ± 12 N) and standard (457 ± 19 N) constructs of 76 mm. Growing rod constructs significantly decreased from 76 mm (522 ± 12 N) to 376 mm active length (200 ± 2 N). Maximum load of growing rods at 76 mm (1084 ± 11 N) was significantly greater than standard at 76 mm (1007 ± 7 N) and growing rods at 376 mm active length (392 ± 5 N). Simulations with active length of 76 mm were within 10% of experimental mechanical characteristics; stress concentrations were at the apex and cranial to connector-rod interaction for standard and growing rod models, respectively.
conclusionGrowing rod constructs are stronger and stiffer than spinal instrumentation constructs; with an increased length accompanied a decrease in strength. Growing rod construct stress concentration locations observed during computational simulation are consistent with clinically observed failure locations. LEVEL OF EVIDENCE: 5.