ArticleJournal of clinical & translational endocrinology2019
Accuracy of five plasma calibrated glucometers to screen for and diagnose gestational diabetes mellitus in a low resource clinic setting.
Article in Journal of clinical & translational endocrinology, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
9 citing papers in PubMed, 14 citations in OpenAlex.
- Implementing a Community-Centered Approach to Gestational Diabetes Screening in Rural Guatemala: A Process Report.Healthcare (Basel, Switzerland) · 2026Article
- Evaluation of the analytical and clinical accuracy of four blood glucose meters in pregnant women with hyperglycaemia.Diabetes, obesity & metabolism · 2025Article
- Performance of point-of-care glucose testing for the diagnosis of gestational diabetes in South Africa.International journal of gynaecology and obstetrics: the official organ of the International Federation of Gynaecology and Obstetrics · 2025Article
- Impact of hematocrit levels on the accuracy of specific blood glucose meters: A hospital-based study.Journal of diabetes investigation · 2024Article
- Article
- Availability and pricing of insulin and related diagnostics in South Africa.Journal of pharmaceutical policy and practice · 2024Article
- Accuracy of capillary blood glucose for 50-g glucose challenge test for gestational diabetes screening.Diabetology international · 2022Article
- Health literacy and diabetes control in pregnant women.Journal of family medicine and primary care · 2020Article
- The impact of differences in plasma glucose between glucose oxidase and hexokinase methods on estimated gestational diabetes mellitus prevalence.Scientific reports · 2019Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
aimsWe investigated the clinical and analytic accuracy of five plasma calibrated glucometers, the use of which is advocated by the World Health Organisation and the International Federation of Gynaecology and Obstetrics, to screen for and diagnose gestational diabetes mellitus (GDM) in low resource settings.
methods592 consecutive black African women underwent a 75 g oral glucose tolerance test (OGTT) at 24-28 weeks gestation at an urban South African community health clinic. Capillary glucose was measured by one of five glucometer brands, each paired with a routine laboratory hexokinase method of plasma glucose measurement. The laboratory results served as the gold standard reference test for GDM diagnosis. World Health Organisation GDM diagnostic thresholds were applied to glucometer and laboratory results.
resultsGlucometer and laboratory determined GDM prevalence was 75/592 (12.7%) and 30/592 (5.1%) with an elevated fasting glucose diagnostic in 64/75 (85%) and 24/30 (80%) of cases respectively. The proportion of glucometer results fulfilling ISO 15197:2013 recommended analytic accuracy at fasting, 60, and 120 min of the OGTT was 92.4%, 49.8% and 61.5%, with Bland Altman method revealing a positive glucometer bias of 0.22 mmol/l (-0.69-1.12 mmol/l), 0.96 mmol/l (-0.65-2.56 mmol/l) and 0.73 mmol/l (-0.73-2.19 mmol/l) respectively. Only three of the glucometer brands evaluated fulfilled ISO 15197:2013 analytic accuracy requirements and this was only achieved at fasting. All glucometers tested were inaccurate at one and two hours of the OGTT.
conclusionsNot all glucometers may be suitable for GDM screening as only three were accurate compared to the reference test and then only at fasting of the OGTT. Importantly, laboratory fasting glucose was diagnostic of GDM in 80% of cases in this study population.