Accuracy of Apple Watch Measurements for Heart Rate and Energy Expenditure in Patients With Cardiovascular Disease: Cross-Sectional Study.

Maarten Falter, Werner Budts, Kaatje Goetschalckx, Véronique Cornelissen, Roselien Buys

Open access · goldAbstract read
In one paragraph

Article in JMIR mHealth and uHealth, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 69 papers, 2 of them syntheses that pooled it.

0numbers the graph read from it
0cells of the map it votes in
69citing papers in PubMed, 2 pooled it
6.3field-weighted citation impact, top 3% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

69 citing papers in PubMed, 2 syntheses or guidelines pooled it, 144 citations in OpenAlex.

  1. Guideline
  2. Pooled it
  3. Trial
  4. Trial
  5. Trial
  6. Article
  7. Review
  8. Article
  9. Article
  10. Article
  11. Article
  12. Article
  13. Review
  14. Article
  15. Article
  16. Article
  17. Review
  18. Article
  19. Observational
  20. Article

9 more citing papers are in PubMed but not listed here.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 1 institution in 1 country.

Maarten FalterCardiology Department, University Hospitals Leuven, Leuven, Belgium.ORCID 0000-0001-7407-115X
Werner BudtsCardiology Department, University Hospitals Leuven, Leuven, Belgium.ORCID 0000-0001-5850-5413
Kaatje GoetschalckxCardiology Department, University Hospitals Leuven, Leuven, Belgium.ORCID 0000-0002-5419-7684
Véronique CornelissenDepartment of Rehabilitation Sciences, KU Leuven, Leuven, Belgium.ORCID 0000-0002-0578-4954
Roselien Buys *Department of Rehabilitation Sciences, KU Leuven, Leuven, Belgium.ORCID 0000-0001-8379-3971
KU Leuven · BE

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundWrist-worn tracking devices such as the Apple Watch are becoming more integrated in health care. However, validation studies of these consumer devices remain scarce.

objectivesThis study aimed to assess if mobile health technology can be used for monitoring home-based exercise in future cardiac rehabilitation programs. The purpose was to determine the accuracy of the Apple Watch in measuring heart rate (HR) and estimating energy expenditure (EE) during a cardiopulmonary exercise test (CPET) in patients with cardiovascular disease.

methodsForty patients (mean age 61.9 [SD 15.2] yrs, 80% male) with cardiovascular disease (70% ischemic, 22.5% valvular, 7.5% other) completed a graded maximal CPET on a cycle ergometer while wearing an Apple Watch. A 12-lead electrocardiogram (ECG) was used to measure HR; indirect calorimetry was used for EE. HR was analyzed at three levels of intensity (seated rest, HR1; moderate intensity, HR2; maximal performance, HR3) for 30 seconds. The EE of the entire test was used. Bias or mean difference (MD), standard deviation of difference (SDD), limits of agreement (LoA), mean absolute error (MAE), mean absolute percentage error (MAPE), and intraclass correlation coefficients (ICCs) were calculated. Bland-Altman plots and scatterplots were constructed.

resultsSDD for HR1, HR2, and HR3 was 12.4, 16.2, and 12.0 bpm, respectively. Bias and LoA (lower, upper LoA) were 3.61 (-20.74, 27.96) for HR1, 0.91 (-30.82, 32.63) for HR2, and -1.82 (-25.27, 21.63) for HR3. MAE was 6.34 for HR1, 7.55 for HR2, and 6.90 for HR3. MAPE was 10.69% for HR1, 9.20% for HR2, and 6.33% for HR3. ICC was 0.729 (P<.001) for HR1, 0.828 (P<.001) for HR2, and 0.958 (P<.001) for HR3. Bland-Altman plots and scatterplots showed good correlation without systematic error when comparing Apple Watch with ECG measurements. SDD for EE was 17.5 kcal. Bias and LoA were 30.47 (-3.80, 64.74). MAE was 30.77; MAPE was 114.72%. ICC for EE was 0.797 (P<.001). The Bland-Altman plot and a scatterplot directly comparing Apple Watch and indirect calorimetry showed systematic bias with an overestimation of EE by the Apple Watch.

conclusionsIn patients with cardiovascular disease, the Apple Watch measures HR with clinically acceptable accuracy during exercise. If confirmed, it might be considered safe to incorporate the Apple Watch in HR-guided training programs in the setting of cardiac rehabilitation. At this moment, however, it is too early to recommend the Apple Watch for cardiac rehabilitation. Also, the Apple Watch systematically overestimates EE in this group of patients. Caution might therefore be warranted when using the Apple Watch for measuring EE.

Indexed as

AgedAged, 80 and overCardiovascular DiseasesCross-Sectional StudiesElectrocardiographyEnergy MetabolismExercise TestFemaleHeart RateHeart Rate DeterminationHumansMaleMiddle AgedMonitoring, PhysiologicWearable Electronic DevicesApple Watchcardiovascular rehabilitationenergy expenditureheart ratemobile healthvalidationwrist-worn devices

Identifiers

PMID30888332
PMCPMC6444219
OpenAlexW2904020811

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.