Evidence map›Paper›PMID 30872166›Full record

ArticleMagnetic resonance imaging2019

Repeatability and reproducibility of longitudinal relaxation rate in 12 small-animal MRI systems.

John C Waterton, Catherine D G Hines, Paul D Hockings, Iina Laitinen, Sabina Ziemian, Simon Campbell, Michael Gottschalk, Claudia Green, Michael Haase, Katja Hassemer and 17 more

Abstract readMulticenter Study
In one paragraph

Article in Magnetic resonance imaging, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 12 papers.

0numbers the graph read from it
0cells of the map it votes in
12citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

12 citing papers in PubMed.

  1. Article
  2. Article
  3. Automated quality control of small animal MR neuroimaging data.Imaging neuroscience (Cambridge, Mass.) · 2024
    Article
  4. Article
  5. Article
  6. Article
  7. Review
  8. Regulation of Oxygen Homeostasis at the Intestinal Epithelial Barrier Site.International journal of molecular sciences · 2021
    Review
  9. Experimental Protocols for MRI Mapping of Renal TMethods in molecular biology (Clifton, N.J.) · 2021
    Article
  10. Article
  11. Article
  12. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

27 authors.

John C WatertonBioxydyn Ltd, Manchester Science Park, Rutherford House, Pencroft Way, MANCHESTER M15 6SZ, United Kingdom; Centre for Imaging Sciences, Division of Informatics Imaging & Data Sciences, School of Health Sciences, Faculty of Biology Medicine & Health, University of Manchester, Manchester Academic Health Sciences Centre, MANCHESTER M13 9PL, United Kingdom. Electronic address: john.waterton@bioxydyn.com.
Catherine D G HinesMerck & Co., Inc., West Point, PA, United States of America. Electronic address: catherine.hines@merck.com.
Paul D HockingsAntaros Medical, BioVenture Hub, 43183 Mölndal, Sweden; MedTech West, Chalmers University of Technology, Gothenburg, Sweden. Electronic address: paul.hockings@antarosmedical.com.
Iina LaitinenSanofi-Aventis Deutschland GmbH, R&D TIM - Bioimaging Germany, Industriepark Höchst, D-65926 Frankfurt am Main, Germany. Electronic address: iina.laitinen@sanofi.com.
Sabina ZiemianBayer AG, Research and Development, Pharmaceuticals, MR and CT Contrast Media Research, Müllerstraße 178, D-13353 Berlin, Germany. Electronic address: sabina.ziemian@bayer.com.
Simon CampbellIn-Vivo Bioimaging UK, RD Platform Technology & Science, GSK Medicines Research Centre, Gunnels Wood Road, STEVENAGE, Hertfordshire, SG1 2NY, United Kingdom. Electronic address: simon.x.campbell@gsk.com.
Michael GottschalkLund University BioImaging Center, Klinikgatan 32, SE-222-42 Lund, Sweden. Electronic address: michael.gottschalk@med.lu.se.
Claudia GreenBayer AG, Research and Development, Pharmaceuticals, MR and CT Contrast Media Research, Müllerstraße 178, D-13353 Berlin, Germany. Electronic address: claudia.green@bayer.com.
Michael HaaseIn-Vivo Bioimaging UK, RD Platform Technology & Science, GSK Medicines Research Centre, Gunnels Wood Road, STEVENAGE, Hertfordshire, SG1 2NY, United Kingdom. Electronic address: michael.v.haase@gsk.com.
Katja HassemerSanofi-Aventis Deutschland GmbH, R&D TIM - Bioimaging Germany, Industriepark Höchst, D-65926 Frankfurt am Main, Germany. Electronic address: katja.hassemer@sanofi.com.
Hans-Paul JuretschkeSanofi-Aventis Deutschland GmbH, R&D TIM - Bioimaging Germany, Industriepark Höchst, D-65926 Frankfurt am Main, Germany.
Sascha KoehlerBruker BioSpin MRI GmbH, Rudolf-Plank-Straße 23, D-76275 Ettlingen, Germany. Electronic address: Sascha.Koehler@bruker.com.
William LloydCentre for Imaging Sciences, Division of Informatics Imaging & Data Sciences, School of Health Sciences, Faculty of Biology Medicine & Health, University of Manchester, Manchester Academic Health Sciences Centre, MANCHESTER M13 9PL, United Kingdom. Electronic address: william.lloyd@manchester.ac.uk.
Yanping LuoiSAT Discovery, Abbvie, 1 North Waukegan Road, North Chicago, IL, 60064-1802, United States of America. Electronic address: yanping.luo@abbvie.com.
Irma Mahmutovic PerssonDepartment of Translational Sciences, Medical Radiation Physics, Lund University, Skåne University Hospital, SE-205 02 Malmö, Sweden. Electronic address: irma.mahmutovic_persson@med.lu.se.
James P B O'ConnorDivision of Cancer Sciences, School of Medical Sciences, Faculty of Biology Medicine & Health, University of Manchester, Manchester Academic Health Sciences Centre, MANCHESTER M20 4BX, United Kingdom. Electronic address: james.o'connor@manchester.ac.uk.
Lars E OlssonDepartment of Translational Sciences, Medical Radiation Physics, Lund University, Skåne University Hospital, SE-205 02 Malmö, Sweden. Electronic address: lars_e.olsson@med.lu.se.
Kashmira PindoriaIn-Vivo Bioimaging UK, RD Platform Technology & Science, GSK Medicines Research Centre, Gunnels Wood Road, STEVENAGE, Hertfordshire, SG1 2NY, United Kingdom. Electronic address: kashmira.2.pindoria@gsk.com.
Jurgen E SchneiderLeeds Institute of Cardiovascular and Metabolic Medicine, University of Leeds, Leeds LS2 9JT, United Kingdom. Electronic address: J.E.Schneider@leeds.ac.uk.
Steven SourbronLeeds Imaging Biomarkers Group, Department of Biomedical Imaging Sciences, University of Leeds, LIGHT Labs, Clarendon Way, LEEDS LS2 9JT, United Kingdom. Electronic address: S.Sourbron@leeds.ac.uk.
Denise SteinmannSanofi-Aventis Deutschland GmbH, R&D TIM - Bioimaging Germany, Industriepark Höchst, D-65926 Frankfurt am Main, Germany. Electronic address: denise.steinmann@sanofi.com.
Klaus StrobelBruker BioSpin MRI GmbH, Rudolf-Plank-Straße 23, D-76275 Ettlingen, Germany. Electronic address: Klaus.Strobel@bruker.com.
Sirisha TadimallaLeeds Imaging Biomarkers Group, Department of Biomedical Imaging Sciences, University of Leeds, LIGHT Labs, Clarendon Way, LEEDS LS2 9JT, United Kingdom. Electronic address: S.Tadimalla@leeds.ac.uk.
Irvin TehLeeds Institute of Cardiovascular and Metabolic Medicine, University of Leeds, Leeds LS2 9JT, United Kingdom. Electronic address: I.Teh@leeds.ac.uk.
Andor VeltienRadboud university medical center, Radiology (766), P.O.Box 9101, 6500, HB, Nijmegen, the Netherlands. Electronic address: Andor.Veltien@radboudumc.nl.
Xiaomeng ZhangiSAT Discovery, Abbvie, 1 North Waukegan Road, North Chicago, IL, 60064-1802, United States of America. Electronic address: xiaomeng.zhang@abbvie.com.
Gunnar SchützBayer AG, Research and Development, Pharmaceuticals, MR and CT Contrast Media Research, Müllerstraße 178, D-13353 Berlin, Germany. Electronic address: gunnar.schuetz@bayer.com.

Funding

Biotechnology and Biological Sciences Research CouncilCancer Research UK 15267Cancer Research UK 22746
6 · The paper itself

Abstract

backgroundMany translational MR biomarkers derive from measurements of the water proton longitudinal relaxation rate R

objectiveTo assess R

methodsR

resultsCoV for day-to-day repeatability (N = 180 regions of interest) was 2.34% and for between-centre reproducibility (N = 9 centres) was 1.43%. Mostly, these do not propagate to biologically significant between-centre error, although a few R

interpretationWhile several factors affect the reproducibility of R

Indexed as

Magnetic Resonance ImagingPhantoms, ImagingAnimalsBiomarkersComputer SimulationImage Processing, Computer-AssistedMiceNickelOxygenProtonsRatsRegression AnalysisReproducibility of ResultsSepharoseWaterBiomarkersNickelOxygenProtonsSepharoseWaterBiomarkerError propagationHardware stabilityMRIPhantomRelaxation timeReproducibility

Identifiers

PMID30872166
PMCPMC6477178

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.