ArticleIEEE transactions on bio-medical engineering2019
Real-Time Visualization of an Acoustically Active Injection Catheter With Ultrasound Imaging: Algorithm and In Vivo Validation in a Swine Model.
Article in IEEE transactions on bio-medical engineering, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed, 3 citations in OpenAlex.
- The current status and evolution of hemodialysis catheters.Renal failure · 2025Review
- A Pilot Study on Novel Use of Color Doppler Imaging for Navigation of Wiring in Coronary Interventions.Ultrasound in medicine & biology · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors at 4 institutions in 1 country.
Funding
Abstract
objectiveTo independently visualize a catheter and needle during minimally invasive surgery in order to aid in precisely guiding them to their intended location.
methodsSymmetric frequency detection allows for the visualization of the acoustically active catheter tip as a unique color in live imaging. This study extends the algorithm to identify two different crystals by unique colors, validating the algorithm with in vivo pig experiments while simulating the human condition using different attenuation pads.
resultsThe catheter and needle tip were identified with unique colors, differentiable from common Doppler colors, with a frame rate varying between 8 and 10 Hz. Both were visible at graded levels of attenuation induced by interposed polymer pads. Reducing ensemble length increased the frame rate and decreased the signal-to-noise ratio (SNR), though not significantly. At the highest in-path attenuation of 12 dB at 5 MHz, the catheter spot marker was visible whereas the needle was not. The SNR of the catheter signal varied between 12.50 and 18.24 dB and the size of the spot marker varied between 149 and 1015 mm
conclusionModified symmetric frequency detection algorithm can uniquely visualize both catheter and needle in real time with in-path attenuation. SIGNIFICANCE: Unambiguous and distinct visualization of separate locations on the catheter facilitates real-time tracking of minimally invasive procedures.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.