Evidence map›Paper›PMID 30843790›Full record

ArticleIEEE transactions on medical imaging2019

Establishment of an Automated Algorithm Utilizing Optical Coherence Tomography and Micro-Computed Tomography Imaging to Reconstruct the 3-D Deformed Stent Geometry.

Mark R Elliott, Dan Kim, David S Molony, Liam Morris, Habib Samady, Sarang Joshi, Lucas H Timmins

Abstract read
In one paragraph

Article in IEEE transactions on medical imaging, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
5citing papers in PubMed, 1 pooled it
0.7field-weighted citation impact, top 27% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

5 citing papers in PubMed, 1 synthesis or guideline pooled it, 8 citations in OpenAlex.

  1. Pooled it
  2. Article
  3. Article
  4. Article
  5. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 3 institutions in 2 countries.

Mark R Elliott
Dan Kim
David S Molony
Liam Morris
Habib Samady
Sarang Joshi
Lucas H Timmins
University of Utah · USEmory University · USGalway-Mayo Institute of Technology · IE

Funding

3D Stent Reconstruction in Bifurcating Lesions By Fusion of OCT and Micro-CT DataR03EB018918 · NIBIB · UNIVERSITY OF UTAH · PI TIMMINS, LUCAS H. · 2015 to 2016
$156k
NIBIB NIH HHS R03 EB018918
6 · The paper itself

Abstract

Percutaneous coronary intervention (PCI) is the prevalent treatment for coronary artery disease, with hundreds of thousands of stents implanted annually. Computational studies have demonstrated the role of biomechanics in the failure of vascular stents, but clinical studies is this area are limited by a lack of understanding of the deployed stent geometry, which is required to accurately model and predict the stent-induced in vivo biomechanical environment. Herein, we present an automated method to reconstruct the 3-D deployed stent configuration through the fusion of optical coherence tomography (OCT) and micro-computed tomography ( μ CT) imaging data. In an experimental setup, OCT and μ CT data were collected in stents deployed in arterial phantoms ( n=4 ). A constrained iterative deformation process directed by diffeomorphic metric mapping was developed to deform μ CT data of a stent wireframe to the OCT-derived sparse point cloud of the deployed stent. Reconstructions of the deployed stents showed excellent agreement with the ground-truth configurations, with the distance between corresponding points on the reconstructed and ground-truth configurations of [Formula: see text]. Finally, reconstructions required <30 min of computational time. In conclusion, the developed and validated reconstruction algorithm provides a complete spatially resolved reconstruction of a deployed vascular stent from commercially available imaging modalities and has the potential, with further development, to provide more accurate computational models to evaluate the in vivo post-stent mechanical environment, as well as clinical visualization of the 3-D stent geometry immediately following PCI.

Indexed as

AlgorithmsStentsCoronary StenosisCoronary VesselsHumansImaging, Three-DimensionalPercutaneous Coronary InterventionPhantoms, ImagingProsthesis FailureTomography, Optical CoherenceX-Ray Microtomography

Identifiers

PMID30843790
PMCPMC6407623
OpenAlexW2892337987

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.